background image

2

K4 Table top controller 4954

KaVo assumes no responsibility for damage
arising through:

• external influences (poor quality of the

media or inadequate installation)

• use of incorrect information
• improper use
• improperly performed repairs.

Repair and maintenance work - apart from
the activities described in these instructions
for use - may be performed only by quali-
fied technical personnel.

In the event of modifications by third par-
ties, the approvals become null and void.
KaVo recommends using only original
spare parts for operation and for repair.

A 1.3 Precautions

Safe operation and protection of the unit are
ensured only through proper use in accor-
dance with the instructions for use and
using the tools approved for the purpose.
The following should also be observed:

• the work safety regulations,
• the accident prevention regulations.

Each time before switching on, check the

set speed.

■ 

Observe the permissible maximum speed

and maximum pressure of the tools
(according to tool manufacturer's
instructions).

A 1 User information

A 1.1 Meaning of the pictograms

Situations where failure to follow the
instructions may lead to danger,

damage to material or operating faults.

Important information for operator
and engineer.

Automatic mode
Automatic sequence

Close, screw in, 
fasten, etc.

Open, release, loosen

+

more, higher

-

less, lower

Continuous operation

Time, time sequence

Disconnect mains plug

A 1.2 Important information

The instructions for use should be
read by the user before starting up

the unit for the first time, in order to avoid
incorrect operation and other damage. If
other language versions are required,
please request these from your responsible
KaVo agent. Duplication and distribution
of the instructions for use (IU) require
KaVo's prior consent.

All technical data, information and proper-
ties of the product described in the IU cor-
respond to the state on going to press.

Modifications and improvements to the
product as a result of new technical devel-
opments are possible.

This does not imply any right to retrofitting
of existing units.

Содержание K4

Страница 1: ...Operating Instructions K4 Table top controller 4954 KaVo Elektrotechnisches Werk GmbH Wangener Stra e 78 D 88299 Leutkirch Tel 0 75 61 86 0 Fax 0 75 61 86 23 3 1 000 3034 WHD RB 11 00 GB 01 33...

Страница 2: ......

Страница 3: ...tential uses 3 A 1 5 Technical data 3 A 1 6 Combination Controller handpiece 3 A 2 Scope of delivery Accessories 4 A 3 Electrical connection 4 A 4 Installation and commissioning 5 A 5 Description 6 A...

Страница 4: ...eed and maximum pressure of the tools according to tool manufacturer s instructions A 1 User information A 1 1 Meaning of the pictograms Situations where failure to follow the instructions may lead to...

Страница 5: ...1000 30 000 min 1 In the event of extreme high frequen cy interfering fields speed variations up to max 20 may occur Rated voltages Voltage ranges 100 V 120 V 230 V 50 60 Hz Output voltage max 30 V D...

Страница 6: ...ltage and rated frequency agree with the data on the rating plate and voltage selector switch If necessary make an adjustment Voltage adjustment By moving the voltage selector switch the country speci...

Страница 7: ...mains plug into the easily accessible correctly installed earthed mains socket This should also be ensured when using an extension cable Insert plug of connecting cable of hand piece in unit socket an...

Страница 8: ...top controller 4954 A 5 Description Malfunction display Speed preselection continuous Foot switch socket Unit socket 6 pin Voltage selector switch with voltage adjustment Mains switch 0 I Foot switch...

Страница 9: ...4 see A 1 6 For correct operation of the handpiece used it is essential to follow the instruc tions for use 1 000 3902 supplied with the handpiece On the K4 controller 4954 press mains switch to posit...

Страница 10: ...which the flash ing sequence begins again In the case of all faults the motor is switched off Some faults are reset when the foot switch is at stop serious errors can be reset only by switching off C...

Страница 11: ...irect current chopper or measuring device U_wr I_wr faulty R Repair motor electronics C Inverter faulty automatic test on swit ching on R Repair motor electronics C Brake device faulty automatic test...

Страница 12: ...HEET FOR USERS K4 TABLE TOP CONTROLLER 4954 10 K4 Table top controller 4954 1 000 1966 1 000 1967 0 200 6528 0 692 9021 0 200 6529 0 245 5015 0 223 2735 0 220 0469 1 000 0906 0 223 4142 Spare parts sh...

Страница 13: ...necting instructions corrosion impurities in the air supply or chemical or electrical influences which are unusual or not admissible in accordance with KaVo s instructions The guarantee shall become n...