background image

8

K4 Table top controller 4954

A 7 Malfunctions

Repair and maintenance work on the
electrical part of the unit may be car-

ried out only by specialists or by persons
who have been trained in the factory and
made aware of the safety regulations.
Before opening housing parts, disconnect
the mains plug or completely isolate the
unit from the mains supply so that no
power reaches the unit.

The individual faults are indicated by
repeating flashes of the green opera-

tion LED @. In the case of fault No. 3, for
example, the LED @ is activated 3 times,
followed by a break, after which the flash-
ing sequence begins again. In the case of
all faults, the motor is switched off. Some
faults are reset when the foot switch is at
stop; serious errors can be reset only by
switching off.

C

• Motor overload, motor current above 

3.4 A for 7 sec

R

• Reduce motor load, stop and start on 

foot switch

C

• Motor blocked for longer than 3 sec

R

• Eliminate blockage, stop and start on

foot switch

C

• No motor or anknown motor connected

R

• Connect K4 motor, stop and start on

foot switch

C

• Open circuit with K 4 motor (1 phase

broken)

R

• Establish connection, stop and start on

foot switch

C

• Mains undervoltage

R

• Check mains voltage and voltage settin-

gon the unit

F

= Fault  

C

= Cause  

R

= Rectification

F

• flash 5 times

F

• flash 4 times 

F

• flash 3 times

F

• flash 2 times 

F

• flash 1 time

@

Содержание K4

Страница 1: ...Operating Instructions K4 Table top controller 4954 KaVo Elektrotechnisches Werk GmbH Wangener Stra e 78 D 88299 Leutkirch Tel 0 75 61 86 0 Fax 0 75 61 86 23 3 1 000 3034 WHD RB 11 00 GB 01 33...

Страница 2: ......

Страница 3: ...tential uses 3 A 1 5 Technical data 3 A 1 6 Combination Controller handpiece 3 A 2 Scope of delivery Accessories 4 A 3 Electrical connection 4 A 4 Installation and commissioning 5 A 5 Description 6 A...

Страница 4: ...eed and maximum pressure of the tools according to tool manufacturer s instructions A 1 User information A 1 1 Meaning of the pictograms Situations where failure to follow the instructions may lead to...

Страница 5: ...1000 30 000 min 1 In the event of extreme high frequen cy interfering fields speed variations up to max 20 may occur Rated voltages Voltage ranges 100 V 120 V 230 V 50 60 Hz Output voltage max 30 V D...

Страница 6: ...ltage and rated frequency agree with the data on the rating plate and voltage selector switch If necessary make an adjustment Voltage adjustment By moving the voltage selector switch the country speci...

Страница 7: ...mains plug into the easily accessible correctly installed earthed mains socket This should also be ensured when using an extension cable Insert plug of connecting cable of hand piece in unit socket an...

Страница 8: ...top controller 4954 A 5 Description Malfunction display Speed preselection continuous Foot switch socket Unit socket 6 pin Voltage selector switch with voltage adjustment Mains switch 0 I Foot switch...

Страница 9: ...4 see A 1 6 For correct operation of the handpiece used it is essential to follow the instruc tions for use 1 000 3902 supplied with the handpiece On the K4 controller 4954 press mains switch to posit...

Страница 10: ...which the flash ing sequence begins again In the case of all faults the motor is switched off Some faults are reset when the foot switch is at stop serious errors can be reset only by switching off C...

Страница 11: ...irect current chopper or measuring device U_wr I_wr faulty R Repair motor electronics C Inverter faulty automatic test on swit ching on R Repair motor electronics C Brake device faulty automatic test...

Страница 12: ...HEET FOR USERS K4 TABLE TOP CONTROLLER 4954 10 K4 Table top controller 4954 1 000 1966 1 000 1967 0 200 6528 0 692 9021 0 200 6529 0 245 5015 0 223 2735 0 220 0469 1 000 0906 0 223 4142 Spare parts sh...

Страница 13: ...necting instructions corrosion impurities in the air supply or chemical or electrical influences which are unusual or not admissible in accordance with KaVo s instructions The guarantee shall become n...

Отзывы: