77.8   S

Storage

Prepared products must be stored, protected from germs (as far as possi‐

ble) and dust, in a dry, dark, cool room.

Comply with the expiry date of the sterilised items.

 

 

 

Setup methods according to DIN EN ISO 17664

55

88   TTools and consumables

Available from dental suppliers.

Material summary

Mat. no.

Instrument stand 2151

0.411.9501

Cellulose pad 100 units

0.411.9862

Cleanpac 10 units

0.411.9691

Nozzle needle

0.410.0931

Sterilisable spray hose (cut section) 0.065.5188

Hose 150 mm (item)

0.593.0252

 

 

 

Tools and consumables

56

O

Only for the USA

Material summary

Mat. no.

KaVo Spray America 2113 A

0.411.9660

QUATTROcare plus Spray America

2141 P

1.005.4524

 

 

Tools and consumables

57

99   TTerms and conditions of warranty

The following warranty conditions apply to this KaVo medical device:

KaVo provides the end customer with a warranty of proper function and

guarantees zero defects in respect of material and processing for a period

of 24 months from data of invoice, subject to the following conditions:

In case of justified complaints, KaVo will honour its warranty with a free

replacement or repair. Other claims of any nature whatsoever, in particular

with respect to compensation, are excluded. In the event of default, gross

negligence or intent, this shall only apply in the absence of mandatory legal

regulations to the contrary.

KaVo cannot be held liable for defects and their consequences that have

arisen or may arise from to natural wear, improper handling, cleaning or

maintenance, non-compliance with operating, maintenance or connection

instructions, calcination or corrosion, contaminated air or water supplies or

chemical or electrical factors deemed abnormal or impermissible in ac‐

 

 

 

Terms and conditions of warranty

58

Содержание INTRAmatic contra-angle 80 ES

Страница 1: ...ger levels 8 2 2 Safety instructions 9 3 Product description 15 3 1 Purpose Proper use 16 3 2 Technical data 18 3 3 Transportation and storage conditions 19 Contents 1 4 4 First use 21 4 1 Check the a...

Страница 2: ...Sterilisation 52 7 8 Storage 55 8 8 Tools and consumables 56 9 Terms and conditions of warranty 58 Contents 4 1 1 U User instructions Dear User Congratulations on purchasing this KaVo quality product...

Страница 3: ...NING WARNING indicates a hazardous situation that can cause death or serious injury Safety 8 DANGER DANGER indicates a hazardous situation that can directly cause death or serious injury 2 2 S Safety...

Страница 4: ...s at KaVo branches throughout the world Technicians specially trained by KaVo The following guidelines must be observed to ensure save use of the elec trically driven handpieces The service instructio...

Страница 5: ...l device may only be used for the described application by a knowledgeable user The following must be observed the applicable health and safety regulations the applicable accident prevention regulatio...

Страница 6: ...RH to 95 RH Product description 19 Air pressure 700 hPa to 1060 hPa 10 psi to 15 psi Protect from moisture Product description 20 4 4 F First use WARNING Hazard from nonsterile products Infection dan...

Страница 7: ...eration 24 5 5 1 A Attaching the medical device Place the medical device on the LUX motor coupling and lock it into place Before each treatment pull on the medical device to see if it is securely seat...

Страница 8: ...r or grinder is securely attached by pulling on it Operation 28 5 5 4 R Removing the milling tool or diamond grinder CAUTION Hazard from rotating tools Laceration Do not carelessly contact rotating to...

Страница 9: ...1 T Troubleshooting Exchanging the O rings on the motor coupling CAUTION Hazard from improper care of the O rings Malfunctions or complete failure of the product Do not use Vaseline or other grease o...

Страница 10: ...fectors in accordance with EN ISO 15883 that are operated with alkaline cleaning agents at a pH of max 10 e g Miele G 7781 G 7881 validation was performed with the VARIO TD programme neodisher medicle...

Страница 11: ...ing agent and extends only to the compatibility of materials with respect to KaVo products For program settings as well as cleansers and disinfectants to be used please refer to the Instructions for U...

Страница 12: ...Z neu tralisation agent and neodisher mielclear rinsing agent and extends only to the compatibility of materials with respect to KaVo products For program settings as well as cleansers and disinfecta...

Страница 13: ...le tip in the opening and spray Carry out the servicing according to the instructions in the section Care with KaVo Spray Setup methods according to DIN EN ISO 17664 48 7 7 5 2 C Care products and sys...

Страница 14: ...KaVo care products Setup methods according to DIN EN ISO 17664 52 CAUTION Contact corrosion due to moisture Damage to product Immediately remove the product from the steam steriliser after the steril...

Страница 15: ...to this KaVo medical device KaVo provides the end customer with a warranty of proper function and guarantees zero defects in respect of material and processing for a period of 24 months from data of i...

Страница 16: ...consequences are derived from manipulations or changes to the product by the customer or a third party Service warranty claims will only be accepted if the product is submitted along with proof of pu...

Отзывы: