CAUTION

Hazard from insufficient spray water.

Insufficient spray water can cause the medical device to overheat and

damage the tooth.

▶ Check the spray water channels and clean the spray nozzles with the

nozzle needle M

Mat. no. 0.410.0931 if necessary.

▶ Check the water filter and exchange if necessary.

First use

23

55   O

Operation

CAUTION

Removing and attaching the contra-angle handpiece while the device is

rotating.

Damage to the drive.

▶ Never attach or remove the contra-angle handpiece while the device

is rotating!

 

Operation

24

55.1   A

Attaching the medical device

Place the medical device on the (LUX) motor coupling and lock it into

place.

Before each treatment, pull on the medical device to see if it is securely

seated in the motor coupling.

 

 

Operation

25

55.2   R

Removing the medical device

Unlock the contra-angle handpiece from the motor coupling and re‐

move it, or pull it off by twisting it slightly.

5.3   IInsert the milling cutters or diamond grinders

Note
Only use carbide cutters or diamond grinders that comply with ISO 1797-1

type 3, are made of steel or hard metal and meet the following criteria:

- Shaft diameter: 1.59 to 1.60 mm

- Overall length: max. 25 mm

- Shaft clamping length: min. 11 mm

- Blade diameter: max. 2 mm

 

 

 

Operation

26

Содержание INTRAmatic contra-angle 80 ES

Страница 1: ...ger levels 8 2 2 Safety instructions 9 3 Product description 15 3 1 Purpose Proper use 16 3 2 Technical data 18 3 3 Transportation and storage conditions 19 Contents 1 4 4 First use 21 4 1 Check the a...

Страница 2: ...Sterilisation 52 7 8 Storage 55 8 8 Tools and consumables 56 9 Terms and conditions of warranty 58 Contents 4 1 1 U User instructions Dear User Congratulations on purchasing this KaVo quality product...

Страница 3: ...NING WARNING indicates a hazardous situation that can cause death or serious injury Safety 8 DANGER DANGER indicates a hazardous situation that can directly cause death or serious injury 2 2 S Safety...

Страница 4: ...s at KaVo branches throughout the world Technicians specially trained by KaVo The following guidelines must be observed to ensure save use of the elec trically driven handpieces The service instructio...

Страница 5: ...l device may only be used for the described application by a knowledgeable user The following must be observed the applicable health and safety regulations the applicable accident prevention regulatio...

Страница 6: ...RH to 95 RH Product description 19 Air pressure 700 hPa to 1060 hPa 10 psi to 15 psi Protect from moisture Product description 20 4 4 F First use WARNING Hazard from nonsterile products Infection dan...

Страница 7: ...eration 24 5 5 1 A Attaching the medical device Place the medical device on the LUX motor coupling and lock it into place Before each treatment pull on the medical device to see if it is securely seat...

Страница 8: ...r or grinder is securely attached by pulling on it Operation 28 5 5 4 R Removing the milling tool or diamond grinder CAUTION Hazard from rotating tools Laceration Do not carelessly contact rotating to...

Страница 9: ...1 T Troubleshooting Exchanging the O rings on the motor coupling CAUTION Hazard from improper care of the O rings Malfunctions or complete failure of the product Do not use Vaseline or other grease o...

Страница 10: ...fectors in accordance with EN ISO 15883 that are operated with alkaline cleaning agents at a pH of max 10 e g Miele G 7781 G 7881 validation was performed with the VARIO TD programme neodisher medicle...

Страница 11: ...ing agent and extends only to the compatibility of materials with respect to KaVo products For program settings as well as cleansers and disinfectants to be used please refer to the Instructions for U...

Страница 12: ...Z neu tralisation agent and neodisher mielclear rinsing agent and extends only to the compatibility of materials with respect to KaVo products For program settings as well as cleansers and disinfecta...

Страница 13: ...le tip in the opening and spray Carry out the servicing according to the instructions in the section Care with KaVo Spray Setup methods according to DIN EN ISO 17664 48 7 7 5 2 C Care products and sys...

Страница 14: ...KaVo care products Setup methods according to DIN EN ISO 17664 52 CAUTION Contact corrosion due to moisture Damage to product Immediately remove the product from the steam steriliser after the steril...

Страница 15: ...to this KaVo medical device KaVo provides the end customer with a warranty of proper function and guarantees zero defects in respect of material and processing for a period of 24 months from data of i...

Страница 16: ...consequences are derived from manipulations or changes to the product by the customer or a third party Service warranty claims will only be accepted if the product is submitted along with proof of pu...

Отзывы: