CAUTION
Hazard from insufficient spray water.
Insufficient spray water can cause the medical device to overheat and
damage the tooth.
▶ Check the spray water channels and clean the spray nozzles with the
nozzle needle M
Mat. no. 0.410.0931 if necessary.
▶ Check the water filter and exchange if necessary.
First use
23
55 O
Operation
CAUTION
Removing and attaching the contra-angle handpiece while the device is
rotating.
Damage to the drive.
▶ Never attach or remove the contra-angle handpiece while the device
is rotating!
Operation
24
55.1 A
Attaching the medical device
▶
Place the medical device on the (LUX) motor coupling and lock it into
place.
▶
Before each treatment, pull on the medical device to see if it is securely
seated in the motor coupling.
Operation
25
55.2 R
Removing the medical device
▶
Unlock the contra-angle handpiece from the motor coupling and re‐
move it, or pull it off by twisting it slightly.
5.3 IInsert the milling cutters or diamond grinders
Note
Only use carbide cutters or diamond grinders that comply with ISO 1797-1
type 3, are made of steel or hard metal and meet the following criteria:
- Shaft diameter: 1.59 to 1.60 mm
- Overall length: max. 25 mm
- Shaft clamping length: min. 11 mm
- Blade diameter: max. 2 mm
Operation
26