![KaVo EXTRAmatic 9040 Скачать руководство пользователя страница 39](http://html2.mh-extra.com/html/kavo/extramatic-9040/extramatic-9040_instructions-for-use-manual_3724524039.webp)
Instructions for use EXTRAmatic 9040
5 Operation | 5.3 Utility programs
▶ Use the "Up" and/or "Down" buttons to set the service life as desired.
▶ Press ENTER.
ð
This saves the new value that has been set.
Factory setting: 800 hours
5.3.12 "Time since carbon filter was changed" utility program (2455)
▶ Using the the up or down button, select the "Carbon filter service life" utility pro‐
gram and choose ENTER.
ð
The total operating time of the active carbon filter is shown.
▶ Press the back button
ð
This takes you out of the utility program.
5.3.13 "Reset active carbon filter monitoring time“ utility program (2550)
▶ Using the the up or down button, select the "Filter time reset" utility program and
choose ENTER.
▶ Use the "Up" and/or "Down" buttons to enter yes or no.
If you select “no“ for filter time reset, the expired time remains unchanged.
If you select “yes“ for filter time reset, the expired time is reset to zero.
▶ Press ENTER.
ð
This saves the new value that has been set.
Choosing "no" will cause you to exit the utility program.
Choosing "yes" will complete the procedure.
▶ Press the back button
ð
This takes you out of the utility program.
5.3.14 "Duration to cut-out" utility program (2600)
▶ Use the "Up" and/or "Down" buttons to start the auxiliary programme, "Overrun
time", and press the "ENTER" button.
▶ Use the "Up" and/or "Down" buttons to set the overrun time as desired (between 3
and 20 seconds).
▶ Press ENTER.
39 / 56
Содержание EXTRAmatic 9040
Страница 1: ...Instructions for use EXTRAmatic 9040 Always be on the safe side...
Страница 55: ......