Instructions for use EXPERTmatic LUX E25 L - 1.007.5550, EXPERTmatic E25 C - 1.007.5551
5 Operation | 5.1 Attaching the medical device
14 / 28
5 Operation
WARNING
Rotating dental bur.
Cuts, infection and burn injury.
4
Never push the press-button while the dental bur is rotating.
4
Do not touch the dental bur while it is rotating.
4
Never touch soft tissue with the handpiece head or instrument cover.
4
Remove the dental bur from the handpiece after treatment to avoid injury
and infection during storage.
CAUTION
Heat transmission during the preparation of abutments.
Thermal damage to the jawbone.
4
During the preparation of abutments, make sure that the preparation times
are short and that there is sufficient cooling.
NOTE
At the beginning of each workday, the water-conducting systems should be
rinsed for at least 2 minutes (without transmission handpieces being at-
tached) and if there is a risk of contamination from reflux or back suction, the
system may also need to be rinsed for 20 to 30 seconds after each patient.
5.1 Attaching the medical device
WARNING
Detaching the medical device during treatment.
A medical device that is not properly locked can detach from the coupling dur-
ing treatment.
4
Before each use, check if the medical device is securely locked onto the
coupling.
CAUTION
Connection to the drive motor.
Straight or contra-angle handpiece jams.
4
Operate the straight or contra-angle handpiece only with the chuck being
closed.
CAUTION
Removing and attaching the straight or contra-angle handpiece while
the drive motor is rotating.
Damage to the catch pin.
4
Never attach or remove the straight or contra-angle handpiece while the
drive motor is rotating.
CAUTION
Pressing the foot switch while attaching or detaching the medical
device.
Property damage to the medical device.
4
Do not connect or remove the medical device while pressing the foot switch.