7 Reprocessing steps in accordance with ISO 17664 | 7.2 Manual Reprocessing

20 / 28

4

Remove the medical device from the spray attachment and let the cleanser
act for 1 minute.

Also refer to:

Instructions for use KaVo CLEANspray

4

If a manual external and internal disinfection do not follow directly, dry the
medical device with KaVo DRYspray.

Also refer to: 

7.2.5 Manual drying, Page 21

7.2.3 Manual external disinfection

 WARNING

Incomplete disinfection.

Infection hazard.

4

Only use disinfection procedures that are verified to be bactericidal, fungi-
cidal and virucidal.

4

If the disinfectants/disinfection procedures fail to meet the mandatory na-
tional requirements, perform a final sterilisation with the sterilisation para-
meters as described.

CAUTION

Never reprocess the medical device with chloride-containing products.

Malfunction and material damage.

4

Reprocess in a washer disinfector or by hand only.

KaVo recommends the following products based on the compatibility of the ma-
terials. The microbiological efficacy must be ensured by the disinfectant manu-
facturer and proven by an expert opinion.

Approved disinfectants:

▪ CaviWipes and CaviCide made by Metrex

▪ Mikrozid AF made by Schülke & Mayr (Liquid or wipes)

▪ FD 322 made by Dürr

Consumables required:

▪ Cloths for wiping the medical device.

4

Spray the disinfectant on a cloth, then wipe down the medical device and al-
low the disinfectant to act according to the instructions of the disinfectant
manufacturer.

4

Follow the instructions for use of the disinfectant.

7.2.4 Manual internal disinfection

 WARNING

Incomplete disinfection.

Infection hazard.

4

Only use disinfection procedures that are verified to be bactericidal, fungi-
cidal and virucidal.

4

If the disinfectants/disinfection procedures fail to meet the mandatory na-
tional requirements, perform a final sterilisation with the sterilisation para-
meters as described.

Содержание EXPERTmatic LUX E25 L

Страница 1: ...Instructions for use EXPERTmatic LUX E25 L 1 007 5550 EXPERTmatic E25 C 1 007 5551...

Страница 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany Phone 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany www kavo com...

Страница 3: ...6 6 Checking for malfunctions and troubleshooting 17 6 1 Checking for malfunctions 17 6 2 Troubleshooting 17 6 2 1 Replacing the O rings on the motor coupling 17 6 2 2 Changing the water filter 18 7 R...

Страница 4: ...ng www kavobox com KaVo Technical Service If you have any questions or complaints please contact the KaVo Technical Service 49 0 7351 56 1000 service instrumente kavo com Target group The instructions...

Страница 5: ...ure range Transportation and storage conditions air pressure Transportation and storage conditions Humidity Protect from moisture Keep dry Protect from impact Original language German Hazard levels Th...

Страница 6: ...Instructions for use EXPERTmatic LUX E25 L 1 007 5550 EXPERTmatic E25 C 1 007 5551 1 User instructions 6 28 CAUTION In cases which if not prevented can lead to material damage...

Страница 7: ...ely before disposal 4 In the case of injury to soft tissue do not continue treatment in the oral cavity with instruments driven by compressed air 4 Use gloves or a finger guard when you test insert an...

Страница 8: ...any damaged motors 4 Do not use motors for unauthorised purposes 2 4 Accessories and combination with other equipment Use of non authorised accessories or non authorised modifications of the device co...

Страница 9: ...ing in house service intervals where the medical device is brought to a professional shop for cleaning servicing and a func tion check Define the service interval depending on the frequency of use As...

Страница 10: ...for the following applications Removal of carious material Removal of fillings Processing of tooth and restoration surfaces Tooth cavity and crown preparations A medical device according to relevant n...

Страница 11: ...ling air flow 5 5 to 9 5 Nl min Pushbutton chuck 1 6 mm Insert Dental burs and diamond grinders in accordance with DIN EN ISO 1797 type 3 Can be attached to All INTRA LUX motors and motors with a conn...

Страница 12: ...g waste management recycling companies Comply with the comprehensive return system KaVo has had its packaging licensed for this purpose Please comply with the regional public waste disposal system CAU...

Страница 13: ...Instructions for use EXPERTmatic LUX E25 L 1 007 5550 EXPERTmatic E25 C 1 007 5551 4 Startup and shut down 4 1 Checking the amount of water 13 28...

Страница 14: ...smission handpieces being at tached and if there is a risk of contamination from reflux or back suction the system may also need to be rinsed for 20 to 30 seconds after each patient 5 1 Attaching the...

Страница 15: ...ith DIN EN ISO 1797 type 3 are made of steel or hard metal and meet the following criteria Shaft diameter 1 59 to 1 60 mm 0 0626 in to 0 0629 in Overall length max 25 mm Shaft clamping length at least...

Страница 16: ...of the dental bur Material damage to dental bur shaft and chuck system reduction of the service life of dental bur and chuck system 4 Do not operate the dental bur at a higher speed than recommended b...

Страница 17: ...ctions for use of motor 6 2 Troubleshooting WARNING Repair WITHOUT using KaVo original spare parts Parts such as the cover can come loose and cause injury Aspiration swallowing of parts and danger of...

Страница 18: ...fection hazard to the dentist and patient 4 Prior to initial startup and after each use reprocess the product and ac cessories CAUTION Overheating of the tooth due to insufficient amount of cooling wa...

Страница 19: ...ves 4 Remove the dental bur from the medical device 4 Remove all residual cement composite or blood immediately 4 Do not place in solutions or similar substances 7 2 Manual Reprocessing WARNING Sharp...

Страница 20: ...ride containing products Malfunction and material damage 4 Reprocess in a washer disinfector or by hand only KaVo recommends the following products based on the compatibility of the ma terials The mic...

Страница 21: ...7 2 5 Manual drying Use KaVo DRYspray for subsequent drying of the air water and gear unit ducts 4 Cover the medical device with the KaVo Cleanpac bag and place it on the corresponding care adapter 4...

Страница 22: ...tor 4 For adaptations unless otherwise specified refer to the Instructions for Use of the washer disinfector 7 3 2 Automated drying The drying procedure is usually part of the cleaning programme of th...

Страница 23: ...e interior in accordance with German Robert Koch Institute requirements KaVo recommends servicing the product as part of the reprocessing after each use i e after each cleaning disinfection and before...

Страница 24: ...n a sterile pack 7 6 Sterilisation Sterilisation in a steam steriliser autoclave in accordance with EN 13060 EN ISO 17665 1 CAUTION Improper service and care Risk of injury 4 Service regularly with su...

Страница 25: ...C 1 007 5551 7 Reprocessing steps in accordance with ISO 17664 7 7 Storage 25 28 7 7 Storage Reprocessed products must be stored appropriately protected from dust in a dry dark and cool space with a l...

Страница 26: ...005 5204 Cleanpac 10 units 0 411 9691 Cellulose pad 100 units 0 411 9862 Nozzle needle 0 410 0921 Replacement filter 1 002 0271 Wrench for water filter 1 002 0321 O ring 8 3x0 68 0 200 6120 KaVo CLEA...

Страница 27: ...aVo shall not be liable for defects and consequences thereof that have arisen or may arise from natural wear improper handling cleaning servicing or main tenance non compliance with operating maintena...

Страница 28: ...1 008 1659 bd 20221209 05 en...

Отзывы: