Instructions for use EXPERTmatic LUX contra-angle E25 L - 1.007.5550 EXPERTmatic contra-angle E25 C - 1.007.5551
2 Safety | 2.2 Safety instructions
CAUTION
Risks due to lack of control equipment.
Hazards can arise if control equipment is not available for changing the speed and
the direction of rotation.
▶ The dental treatment unit connected must have control equipment for changing
the speed and direction of rotation.
▶ A note is to be included in the documents accompanying the dental treatment
unit, referring to responsibilities arising from safety, reliability and performance.
▶ The medical device may only be combined with a treatment centre released by
KaVo.
CAUTION
Risk due to incorrectly stored handpiece.
Injury and infection caused by chucked cutter or grinder.
▶ After treatment, place the handpiece properly in the cradle, without the cutter or
grinder.
CAUTION
Risk of burn injury from hot handpiece head or hot handpiece lid.
Burn injuries in the mouth may be caused if the handpiece overheats.
▶ Never touch soft tissue with the handpiece head or handpiece lid!
CAUTION
Premature wear and malfunctioning from improper storage during long periods of
nonuse.
Reduced product life.
▶ The medical device should be cleaned, serviced and stored in a dry location, ac‐
cording to instructions, before long periods of nonuse.
CAUTION
Hazard from use as a light probe.
Do not use the medical device as a light probe.
▶ Use an appropriate light probe for additional illumination of the oral cavity or site
of preparation.
CAUTION
Hazard from the use of handpieces equipped with electronic micromotors.
Electronic micromotors generate much more energy than conventional pneumatic tur‐
bines and motors. Given the higher torque and speed, handpieces that are poorly
serviced, damaged or used improperly can overheat which can cause serious burn in‐
juries to the patient.
▶ Observe the following points.
Note
For safety reasons, we recommend that the tool holder system be checked annually
after the warranty period expires.
The following guidelines must be observed to ensure save use of the electrically driv‐
en handpieces:
▪ The service instructions for handpieces must be precisely following when using
KAVO spray or QUATTROcare care systems.
▪ Before each use, the handpiece must be checked for external damage.