Instructions for use EXPERTmatic LUX contra-angle E25 L - 1.007.5550 EXPERTmatic contra-angle E25 C - 1.007.5551

2 Safety | 2.2 Safety instructions

 CAUTION

Risks due to lack of control equipment.
Hazards can arise if control equipment is not available for changing the speed and

the direction of rotation.

▶ The dental treatment unit connected must have control equipment for changing

the speed and direction of rotation.

▶ A note is to be included in the documents accompanying the dental treatment

unit, referring to responsibilities arising from safety, reliability and performance.

▶ The medical device may only be combined with a treatment centre released by

KaVo.

 CAUTION

Risk due to incorrectly stored handpiece.
Injury and infection caused by chucked cutter or grinder.

▶ After treatment, place the handpiece properly in the cradle, without the cutter or

grinder.

 CAUTION

Risk of burn injury from hot handpiece head or hot handpiece lid.
Burn injuries in the mouth may be caused if the handpiece overheats.

▶ Never touch soft tissue with the handpiece head or handpiece lid!

 CAUTION

Premature wear and malfunctioning from improper storage during long periods of

nonuse.
Reduced product life.

▶ The medical device should be cleaned, serviced and stored in a dry location, ac‐

cording to instructions, before long periods of nonuse.

 CAUTION

Hazard from use as a light probe.
Do not use the medical device as a light probe.

▶ Use an appropriate light probe for additional illumination of the oral cavity or site

of preparation.

 CAUTION

Hazard from the use of handpieces equipped with electronic micromotors.
Electronic micromotors generate much more energy than conventional pneumatic tur‐

bines and motors. Given the higher torque and speed, handpieces that are poorly

serviced, damaged or used improperly can overheat which can cause serious burn in‐

juries to the patient.

▶ Observe the following points.

Note

For safety reasons, we recommend that the tool holder system be checked annually

after the warranty period expires.

The following guidelines must be observed to ensure save use of the electrically driv‐

en handpieces:

▪ The service instructions for handpieces must be precisely following when using

KAVO spray or QUATTROcare care systems.

▪ Before each use, the handpiece must be checked for external damage.

 

7 / 26

Содержание EXPERTmatic LUX contra-angle E25 L

Страница 1: ...Instructions for use EXPERTmatic LUX contra angle E25 L 1 007 5550 EXPERTmatic contra angle E25 C 1 007 5551 Always be on the safe side...

Страница 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Phone 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Страница 3: ...the O rings on the motor coupling 14 6 2 2 Changing the water filter 14 7 Setup methods according to DIN EN ISO 17664 16 7 1 Preparation at the site of use 16 7 2 Cleaning 16 7 2 1 Manual cleaning ext...

Страница 4: ...9 Terms and conditions of warranty 23 Instructions for use EXPERTmatic LUX contra angle E25 L 1 007 5550 EXPERTmatic contra angle E25 C 1 007 5551 Content 4 26...

Страница 5: ...pyright by KaVo Dental GmbH Symbols Refer to the chapter on Safety Warning symbol Important information for users and service technicians Can be steam sterilised at 134 o C 1 o C 4 o C 273 o F 1 6 o F...

Страница 6: ...nger levels The safety instructions listed here together with the three levels of danger will help avert property damage and injury CAUTION CAUTION indicates a hazardous situation that can cause damag...

Страница 7: ...d or handpiece lid CAUTION Premature wear and malfunctioning from improper storage during long periods of nonuse Reduced product life The medical device should be cleaned serviced and stored in a dry...

Страница 8: ...ervice KaVo products Technicians at KaVo branches throughout the world Technicians specially trained by KaVo To ensure proper function the medical device must be set up according to the reproc essing...

Страница 9: ...r the following applications Removal of carious material cavity and crown prep arations removal of fillings processing of tooth and restoration surfaces A medical device according to relevant national...

Страница 10: ...ra angle handpiece can be mounted on all INTRAmatic LUX motors and motors with a connector in accordance with ISO 3964 DIN 13940 3 3 Transportation and storage conditions CAUTION It is hazardous to st...

Страница 11: ...oist cooling air can cause malfunctions and lead to premature bearing wear Make sure that the supply of cooling air is dry clean and uncontaminated accord ing to ISO 7494 2 4 1 Check the amount of wat...

Страница 12: ...can damage the medical device and motor coupling Do not connect or remove the medical device while pressing the footswitch Lightly spray O rings on motor coupling with Mehrfunktionsspritze Spray Attac...

Страница 13: ...o prevent an injury or infection Press the push button firmly with your thumb and simultaneously insert the milling or grinding tool until it contacts the stop Check that the cutter or grinder is secu...

Страница 14: ...for the medical device An O ring is missing on the motor coupling Replace O ring 6 2 Troubleshooting 6 2 1 Exchanging the O rings on the motor coupling CAUTION Hazard from improper care of the O rings...

Страница 15: ...ngle E25 L 1 007 5550 EXPERTmatic contra angle E25 C 1 007 5551 6 Troubleshooting 6 2 Troubleshooting Screw out and remove the filter with the wrench Mat no 1 002 0321 Insert the new filter Mat no 1 0...

Страница 16: ...in a thermodisinfector 7 2 1 Manual cleaning external Accessories required Tap water 30 o C 5 o C 86 o F 10 o F Brush e g medium hard toothbrush Brush off under flowing tap water 7 2 2 Automated exte...

Страница 17: ...nternal cleaning KaVo recommends thermodisinfectors in accordance with EN ISO 15883 1 which are operated with alkaline cleaning agents with a pH value of max 10 e g Miele G 7781 G 7881 Validation was...

Страница 18: ...in accordance with EN ISO 15883 1 which are operated with alkaline cleaning agents with a pH value of max 10 e g Miele G 7781 G 7881 Validation was carried out with Programme VARIO TD cleaning agent n...

Страница 19: ...ress the spray button for one second Servicing of the clamping chuck KaVo recommends cleaning and servicing the chuck system once a week Remove the cutter or grinder place the spray nipple tip in the...

Страница 20: ...able for the sterilisation procedure Weld each medical device individually in a sterilised item package 7 7 Sterilisation Sterilisation in a steam steriliser autoclave in accordance with EN 13060 ISO...

Страница 21: ...007 5551 7 Setup methods according to DIN EN ISO 17664 7 8 Storage Use according to the manufacturer s Instructions for Use 7 8 Storage Prepared products must be stored protected from germs as far as...

Страница 22: ...ers Material summary Mat no Instrument stand 2151 0 411 9501 Cellulose pad 100 units 0 411 9862 Cleanpac 10 units 0 411 9691 Nozzle needle 0 410 0921 Replacement filter 1 002 0271 Key 1 002 0321 O rin...

Страница 23: ...not be liable for defects and their consequences that have arisen or may arise from natural wear improper handling cleaning or maintenance non compliance with operating maintenance or connection inst...

Страница 24: ...Instructions for use EXPERTmatic LUX contra angle E25 L 1 007 5550 EXPERTmatic contra angle E25 C 1 007 5551 9 Terms and conditions of warranty 7 8 Storage 24 26...

Страница 25: ......

Страница 26: ...1 008 4099 kb 20110218 01 en...

Отзывы: