Instructions for use EXPERTmatic LUX contra-angle E25 L - 1.007.5550 EXPERTmatic contra-angle E25 C - 1.007.5551
7 Setup methods according to DIN EN ISO 17664 | 7.4 Drying
18 / 26
Consumables required:
▪ Cloths for wiping off the medical device.
▶ Spray the disinfectant on a cloth, then thoroughly wipe down the medical device
and leave the disinfectant to soak in according to the instructions from the disin‐
fectant manufacturer.
▶ Follow the instructions for use of the disinfectant.
7.3.2 Manual disinfection - internal
The efficacy of manual internal disinfection must be demonstrated by the manufactur‐
er of the disinfection agent. With KaVo products, use only disinfection agents that
have been released by KaVo with respect to the compatibility of materials (e.g. WL-
cid / made by ALPRO).
▶ Immediately after internal disinfection, lubricate the KaVo medical device immedi‐
ately with care agents from the KaVo care system.
▶ Follow the instructions for use of the disinfectant.
7.3.3 Machine disinfection - external and internal
KaVo recommends thermodisinfectors in accordance with EN ISO 15883-1, which are
operated with alkaline cleaning agents with a pH value of max. 10 (e.g. Miele G
7781 / G 7881 – Validation was carried out with Programme "VARIO-TD", cleaning
agent "neodisher® mediclean", neutralisation agent "neodisher® Z" and rinsing agent
"neodisher® mielclear" and only applies to the material compatibility with KaVo prod‐
ucts).
▶ For program settings as well as cleansers and disinfectants to be used, please re‐
fer to the Instructions for Use of the thermodisinfector (complying with max. pH
value of 10).
▶ In order to prevent negative effects on the medical device, make sure that the in‐
terior and the exterior of the medical device are dry after completion of the cycle,
and then grease it immediately with servicing agents from the KaVo care system.
7.4 Drying
Manual Drying
▶ Blow off the outside and inside with compressed air until water drops are no lon‐
ger visible.
Automatic Drying
The drying procedure is normally part of the cleaning program of the thermodisinfec‐
tor.
▶ Follow the instructions for use of the thermodisinfector.
7.5 Care products and systems - Servicing
WARNING
Sharp cutters or grinders in the medical device.
Risk of injury from sharp and/or pointed cutters or grinders.
▶ Remove cutter or grinder.