Instructions for use EXPERTmatic LUX contra-angle E25 L - 1.007.5550 EXPERTmatic contra-angle E25 C - 1.007.5551
7 Setup methods according to DIN EN ISO 17664 | 7.5 Care products and systems - Servicing
CAUTION
Premature wear and malfunctions from improper servicing and care.
Reduced product life.
▶ Perform proper care regularly!
Note
KaVo only guarantees that its products will function properly when the care prod‐
ucts are used that are listed as accessories since they were tested for proper use
on our products.
7.5.1 Care with KaVo Spray
▶ Remove the cutter or grinder.
▶ Cover the product with the Cleanpac bag.
▶ Plug the product onto the cannula, and press the spray button for one second.
Servicing of the clamping chuck
KaVo recommends cleaning and servicing the chuck system once a week.
▶ Remove the cutter or grinder, place the spray nipple tip in the opening and spray.
▶ Carry out the servicing according to the instructions in the section, "Care with Ka‐
Vo Spray".
7.5.2 Care with KaVo SPRAYrotor
▶ Place the product on the appropriate coupling of the KaVo SPRAYrotor and cover
it with a Cleanpac bag.
▶ Servicing the product.
See also:
2
Instructions for use KaVo SPRAYrotor
7.5.3 Servicing with KaVo QUATTROcare
Cleaning and servicing unit with expansion pressure for effective cleaning and care.
KaVo recommends servicing the product as part of the reprocessing after each use,
i.e. after each cleaning, disinfection, and before each sterilisation.
▶ Remove the cutter or grinder.
▶ Servicing the product.
19 / 26