Instructions for use DIAGNOcam 2170 U
9 Information concerning the electromagnetic compatibility according to EN IEC 60601-1-2 | 9.4 Recommended
safe distance between portable and mobile HF telecommunications equipment and the treatment unit
characteristics are observed, additional measures may be required, such as,
e.g., changing the orientation or using a different location for the DIAGNOcam
2170 U.
d
In the frequency range of 150 kHz to 80 MHz, the field strength should be less
than 3V
eff
V/m.
9.4 Recommended safe distance between portable and
mobile HF telecommunications equipment and the
treatment unit
The DIAGNOcam 2170 U is intended for use in an electromagnetic environment
in which the HF interference parameters are controlled. The customer or the
user of the DIAGNOcam 2170 U can help prevent electromagnetic interference
by maintaining the minimum distance between portable and mobile HF tele-
communications devices (transmitters) and the DIAGNOcam 2170 U depending
on the output of the communication device as indicated below.
Safe distance depending on the transmission frequency:
Rated maximum out-
put power of trans-
mitter in W
150 kHz to 80 MHz
d=1.17
P
m
80 MHz to 800 MHz
d=1.17
P
m
800 MHz to 2.5 GHz
d=2.33
P
m
0.01
0.1
0.1
0.2
0.1
0.4
0.4
0.7
1
1.2
1.2
2.3
10
3.7
3.7
7.4
100
11.7
11.7
23.3
For transmitters whose maximum rated power is not included in the above
table, the recommended safe distance d in metres (m) can be calculated using
the equation for the respective column, where P is the maximum rated power
of the transmitter in Watts (W) as specified by the manufacturer.
NOTE: At 80 MHz and 800 MHz, the higher frequency range applies.
The spread of electromagnetic waves is absorbed and reflected by buildings,