Instructions for use SMARTmatic PROPHY S19 - 1.011.6740 SMARTmatic PROPHY S19 K - 1.011.6741
7 Reprocessing steps in accordance with ISO 17664 | 7.7 Sterilisation
▶ Seal the medical device individually in the sterilised item packaging.
7.7 Sterilisation
Sterilisation in a steam steriliser (autoclave) in accordance with EN
13060 / ISO 17665-1
NOTICE
Improper service and care.
Premature wear and reduced product service life.
▶ Before each sterilisation cycle, service the medical device with KaVo care
products.
NOTICE
Contact corrosion due to moisture.
Damage to product.
▶ Immediately remove the product from the steam steriliser after the sterilisation
cycle.
The KaVo medical device has a maximum temperature resistance up to 138 ℃
(280.4 °F).
Select a suitable procedure (depending on the available autoclave) from the following
sterilisation processes:
▪ Autoclave with three times pre-vacuum:
- at least 3 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4 °F)
▪ Steriliser using the gravity method:
- at least 10 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4 °F)
▶ Use according to the manufacturer's Instructions for Use.
7.8 Storage
Prepared products must be stored, protected from germs (as far as possible) and
dust, in a dry, dark, cool room.
Note
Comply with the expiry date of the sterilised items.
23 / 28