Instructions for use SMARTmatic PROPHY S19 - 1.011.6740 SMARTmatic PROPHY S19 K - 1.011.6741

7 Reprocessing steps in accordance with ISO 17664 | 7.4 Drying

Consumables required:

▪ Cloths for wiping the medical device.

▶ Spray the disinfectant on a cloth, then wipe down the medical device and allow

the disinfectant to act according to the instructions of the
disinfectant manufacturer.

Note

Follow the instructions for use of the disinfectant.

7.3.2 Manual internal disinfection

The efficacy of manual internal disinfection must be demonstrated by the manufac-
turer of the disinfection agent. With KaVo products, use only disinfection agents that
have been released by KaVo with respect to the compatibility of materials (e.g. WL-
cid / made by ALPRO).

▶ Blow off with compressed air until no water drops are visible.

▶ Immediately after internal disinfection, lubricate the KaVo medical device with care

agents from the KaVo care system.

7.3.3 Automated external and internal disinfection

KaVo recommends washer disinfectors in compliance with EN ISO 15883-1 that are
operated with alkaline cleaning agents.
The validations were conducted with the VARIO-TD program, the cleaning agent
neodisher® MediClean and the neutralisation agent neodisher® Z.

▶ For programme settings as well as cleansers and disinfectants to be used, please

refer to the Instructions for Use of the washer disinfector.

▶ In order to prevent negative effects on the KaVo medical device, make sure that

the interior and the exterior of the medical device are dry after completion of the
cycle, and then grease it immediately with servicing agents from the KaVo care
system.

7.4 Drying

Manual Drying

▶ Blow off the outside and inside with compressed air until water drops are no

longer visible.

Automatic Drying

The drying procedure is usually part of the cleaning programme of the washer disin-
fector.

▶ Follow the instructions for use of the thermodisinfector.

7.5 Care products and systems - Servicing

NOTICE

Improper service and care.
Premature wear and reduced product service life.

▶ Service regularly with suitable agents.

19 / 28

Содержание 1.011.6740

Страница 1: ...Instructions for use SMARTmatic PROPHY S19 1 011 6740 SMARTmatic PROPHY S19 K 1 011 6741...

Страница 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Phone 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Страница 3: ...ngs 16 7 Reprocessing steps in accordance with ISO 17664 17 7 1 Preparations at the site of use 17 7 2 Cleaning 17 7 2 1 Manual external cleaning 17 7 2 2 Automated external cleaning 17 7 2 3 Manual i...

Страница 4: ...avo com Target group This document is intended for dentists and their assistants The startup section is also intended for service technicians General marks and symbols Refer to the chapter on Safety W...

Страница 5: ...PROPHY S19 1 011 6740 SMARTmatic PROPHY S19 K 1 011 6741 1 User instructions CAUTION In cases which if not prevented could lead to minor or moderate injury NOTICE In cases which if not prevented coul...

Страница 6: ...In the case of injury to soft tissue do not continue treatment in the oral cavity with instruments driven by compressed air Use gloves or finger guard whenever you test insert and remove the tool 2 2...

Страница 7: ...acility for changing the speed and the direction of rotation must be present The medical device may only be combined with a treatment centre released by KaVo Comply with the Instructions for Use of th...

Страница 8: ...19 K 1 011 6741 2 Safety 2 5 Service and repair 8 28 KaVo recommends specifying in house service intervals where the medical device is brought to a professional shop for cleaning servicing and a funct...

Страница 9: ...ool to support the periodontal treatment by polish ing with a rotating tool A medical device according to relevant national statutory regulations Proper use According to these provisions this medical...

Страница 10: ...mission 8 1 Cooling air flow 5 5 to 9 5 Nl min Prophylaxis heads with a Doriot connector can be inserted Shaft and or drive shaft in accordance with DIN EN ISO 14457 for metallic and ceramic materials...

Страница 11: ...Transportation and storage conditions NOTICE Startup after refrigerated storage Malfunction Prior to startup very cold products must be heated up to a temperature of 20 C to 25 C 68 F to 77 F Temperat...

Страница 12: ...Dispose of the product in appropriate manner Infection hazard Before disposal reprocess and sterilise the product and accessories appropri ately NOTICE Damage from soiled and moist cooling air Contami...

Страница 13: ...to ensure that it is se curely locked onto the motor coupling NOTICE Removing and attaching the handpiece while the drive motor is rotating Damage to the driver Never attach or remove the handpiece w...

Страница 14: ...is heads Injury to the patient or damage to the medical device Comply with the instructions for use and the intended use of the tools Only use prophylaxis heads that do not deviate from the specified...

Страница 15: ...the prophylaxis head 5 4 Removing the prophylaxis head Remove the prophylaxis head axially from the handpiece CAUTION Hazard due to incorrectly stored instrument Injury and infection from clamped prop...

Страница 16: ...is irregularly heated The medical device is too hot while working Service the medical device When the speed drops or is uneven Service the medical device An O ring is missing on the motor coupling Re...

Страница 17: ...lly or in a washer disinfector only 7 2 1 Manual external cleaning Accessories required Tap water 30 o C 5 o C 86 o F 10 o F Brush e g medium hard toothbrush Brush off under flowing tap water 7 2 2 Au...

Страница 18: ...line cleaning agents The validations were conducted with the VARIO TD program the cleaning agent neodisher MediClean and the neutralisation agent neodisher Z For program settings as well as cleansers...

Страница 19: ...external and internal disinfection KaVo recommends washer disinfectors in compliance with EN ISO 15883 1 that are operated with alkaline cleaning agents The validations were conducted with the VARIO T...

Страница 20: ...rvicing with KaVo Spray KaVo recommends servicing the product as part of the reprocessing after each use i e after each cleaning disinfection and before each sterilisation Remove the tool Cover the pr...

Страница 21: ...xpansion pressure for the interior cleaning of inor ganic residues and optimum care no validated cleaning of the interior in accordance with German RKI requirements KaVo recommends servicing the produ...

Страница 22: ...rvice button for at least three seconds until the spray canister control LED flashes three times consecutively The device is in chuck service mode Remove the service coupling of the chuck from the sid...

Страница 23: ...to moisture Damage to product Immediately remove the product from the steam steriliser after the sterilisation cycle The KaVo medical device has a maximum temperature resistance up to 138 280 4 F Sele...

Страница 24: ...ent stand 2151 0 411 9501 Cleanpac 10 units 0 411 9691 Cellulose pad 100 units 0 411 9862 Spray head INTRA KaVo Spray 0 411 9911 INTRA service coupling 1 009 6143 Material summary Mat No KaVo Spray 21...

Страница 25: ...for defects and their consequences that have arisen or may arise from natural wear improper handling cleaning or maintenance non compliance with operating maintenance or connection instructions calcin...

Страница 26: ......

Страница 27: ......

Страница 28: ...1 010 0216 s 20170120 2 en...

Отзывы: