10 Troubleshooting
57 / 68
Malfunction
Cause
Remedy
Insufficient coolant flow in the
instrument.
Spray nozzles are crusty
or soiled.
4
Clean the spray nozzles with the nozzle
needle or reprocess the part.
Also refer to:
Instructions for use
SURGmatic
Glass bottle containing
the coolant is not ventil-
ated.
4
Open the cap on the ventilation valve of
the puncture needle.
The motor makes a grinding
noise or does not run
smoothly.
The motor is not cor-
rectly plugged on or
screwed on.
4
Firmly insert the motor hose into the
housing.
4
Firmly screw on the motor hose to the
motor.
4
Check if all the connections and couplings
are firmly seated.
No light on the straight and
contra-angle handpiece.
The light is not turned
on.
4
Turn on the light.
Also refer to:
5.2 Make the device settings, Page 29
The straight and contra-
angle handpiece is im-
properly attached.
4
Attach the straight and contra-angle
handpiece until the guide stud audibly
locks.
Defective LED.
4
Replace the LED.
Also refer to:
Instructions for use INTRA
LUX S600 LED
Not a suitable straight
and contra-angle hand-
piece.
4
Use a suitable light, straight and contra-
angle handpiece.
Error message from software
Malfunction
Cause
Remedy
Event E3:
Release foot pedal
Foot control was activ-
ated while the unit star-
ted up.
4
Release the foot control.
Event E4:
Data initialisation
Setting data newly ini-
tialised.
4
Confirm message and check or correct the
programme settings.
4
If the problem persists, notify the service.
Event E6:
Internal communication error
Internal system error.
4
Turn the unit off and on. If the problem
persists, notify service engineer.
Event E9:
Set time and date
Time was not reset after
re-start.
4
Setting the date and time.
4
If the problem persists, have a service en-
gineer replace the battery of the real-time
clock.
Events E29:
SD card write protection
SD memory card is
write-protected or de-
fective.
4
Switch the write-protect switch on the left
side of the card upwards to "unlock",
"write" or the like.
4
If the problem persists, use a new SD
memory card.
4
If the problem persists, notify service en-
gineer.
All events > 30:
Hardware error
Internal system error.
4
Turn the unit off and on. If the problem
persists, notify service engineer.
Event E36 and E37:
Motor overload
Overtemperature on
stator.
Over-current > nominal
current.
4
Let the motor rest. If the problem per-
sists, notify service engineer.