Instructions for use PROPHYflex 3 - 2018 - 1.000.4672, - 1.006.9926, - 1.006.9928
7 Preparation methods according to ISO 17664 | 7.7 Sterilisation
22 / 26
▶ Individually seal the medical device in the sterilised item packaging.
7.7 Sterilisation
Sterilisation in a steam steriliser (autoclave) in accordance with EN
13060 / ISO 17665-1
CAUTION
Contact corrosion due to moisture.
Damage to product.
▶ Immediately remove the product from the steam steriliser after the sterilisation cy‐
cle!
Note
Prior to attaching the powder container, all powder-conducting parts and air chan‐
nels must be absolutely dry. Screw together the powder container and handpiece
only in the cold state.
Note
Unscrew the powder container anticlockwise and drain and clean it before each
thermal disinfection or sterilisation. Do not screw the powder container back on be‐
fore thermal disinfection or sterilisation. Likewise, clean off any powder residue from
the PROPHYflex, especially the cannulas, tubes and P-nozzle.
The KaVo medical device has a maximum temperature resistance of up to 138 ℃
(280.4 °F).
Select a suitable procedure (depending on the available autoclave) from the following
sterilisation processes:
▪ Autoclave with three times pre-vacuum:
- at least 3 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4 °F)
▪ Autoclave using the gravity method:
- at least 10 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4 °F)
▶ Use according to the manufacturer's Instructions for Use.
7.8 Storage
▶ Reprocessed products should be stored protected from dust with minimum expo‐
sure to germs in a dry, dark and cool space.
▶ Comply with the expiry date of the sterilised items.
Содержание 1.000.4672
Страница 1: ...Instructions for use PROPHYflex 3 2018 1 000 4672 1 006 9926 1 006 9928 Always be on the safe side ...
Страница 25: ......
Страница 26: ...1 000 6601 kb 20140210 07 en ...