Instructions for use PROPHYflex 3 - 2018 - 1.000.4672, - 1.006.9926, - 1.006.9928
7 Preparation methods according to ISO 17664 | 7.3 Disinfection
20 / 26
▶ Cover the medical device with the KaVo CLEANpac bag, and place it on the cor‐
responding care adapter. Press the spray button three times for 2 seconds each
time. Remove the medical device from the spray attachment and let the cleaner
work for one minute.
▶ Afterwards, rinse for 3-5 seconds with KaVo DRYspray.
See also:
2
KaVo CLEANspray / KaVo DRYspray Instructions for Use
Note
KaVo CLEANspray and KaVo DRYspray for manual interior cleaning are only avail‐
able in the following countries:
Germany, Austria, Switzerland, Italy, Spain, Portugal, France, Luxembourg, Bel‐
gium, Netherlands, United Kingdom, Denmark, Sweden, Finland and Norway.
In other countries interior cleaning can only be carried out with thermodisinfectors in
accordance with EN ISO 15883-1.
7.2.4 Automated internal cleaning
KaVo recommends thermodisinfectors in accordance with EN ISO 15883-1, which are
operated with alkaline cleaning agents with a pH value of max. 10 (e.g. Miele G
7781/G 7881 – Validation was carried out with Programme "VARIO-TD", cleaning
agent "neodisher® mediclean", neutralisation agent "neodisher® Z" and rinsing agent
"neodisher® mielclear" and only applies to the material compatibility with KaVo prod‐
ucts).
▶ For program settings as well as cleansers and disinfectants to be used, please re‐
fer to the Instructions for Use of the thermodisinfector (complying with max. pH
value of 10).
7.3 Disinfection
CAUTION
Malfunctioning from using a disinfectant bath or disinfectant containing chlorine.
Defects in the product.
▶ Only disinfect in a thermodisinfector or manually.
7.3.1 Manual external disinfection
KaVo recommends the following products based on compatibility of the materials. The
microbiological efficacy must be ensured by the disinfectant manufacturer.
▪ Mikrozid AF made by Schülke & Mayr (liquid or cloths)
▪ FD 322 made by Dürr
▪ CaviCide made by Metrex
Consumables required:
▪ Cloths for wiping off the medical device.
▶ Spray the disinfectant on a cloth, then thoroughly wipe down the medical device
and leave the disinfectant to soak in according to the instructions from the disin‐
fectant manufacturer.
▶ Follow the instructions for use of the disinfectant.