background image

16

 

Cod. 2050-0001  Rev 1.1 

5.4

 

PRELOAD OF THE EXPANSION TANK 

 

For the preload of the expansion tank follow these steps: 

 
Disconnect the power supply 

Empty the expansion tank completely supplying water. 

With the use of a compressor and a tire gun, bring the tank preload pressure to 0.1 MPa. 

Put the machine in the original position and connect the power plug. 

 

 

5.5

 

EQUIPMENT SANIFICATION 

 

Run the machine sanitation before each filter change, and in any case at intervals not 
exceeding 6 months. 
Run also sanitizes after prolonged disuse of the machine and at the first starting. 
All sanitation operations must be performed by qualified personal. 
Use extreme caution in the use and dosage of chemicals. 
Use protective clothing to the skin, hands and eyes as shown in the safety data sheet 
of chemicals used. 

Close the mix valve on the front of the machine and fill a tank with 10 liters of osmotic 
water. 

Add hydrogen peroxide 10 vol (3%) to achieve a concentration of 0.2% by weight: 
dose 650 ml of hydrogen peroxide at 10 volumes per 10 liters of osmotic water or 
dose 250 ml of hydrogen peroxide at 24 volumes per 10 liters of osmotic water or 
dose 180 ml of hydrogen peroxide at 36 volumes per 10 liters of osmotic water or 
dose 50 ml of hydrogen peroxide at 130 volumes per 10 liters of osmotic water. 
Close the water intake valve and disconnect the water intake pipes, drain and permeated 
by the machinery, put a bridge on the low pressure switch. 
Connect the water pipes of entry, drain and permeated by the machine at the tank. 
Slightly open the mixing valve, turn the machine on and recirculate the sanitizing solution 
for at least 10 minutes. 
Turn off the machine and wait 10 minutes until the solution act. 
Reconnect the pipes of drain and permeate, previously unplugged by the machine. 
Turn on the machine and deliver up to empty the tank of the sanitizing solution. 
Replace the filter and reconnect the minimum pressure switch. 
Turn on the machine and deliver water for at least 10 minutes to drain any residual 
sanitizing solution from the machine. 

In the event of system resting up, carry out the standard sanitation cycle § 5.5. 
Disconnect hydraulically the system and keep it away from frost or excess heat, 
every 6 months of stopping you must carry out a new disinfection cycle. 

Содержание EASY:FAST

Страница 1: ...1 Cod 2050 0001 Rev 1 1 EASY FAST MANUAL EQUIPMENT FOR POTABLE WATER TREATMENT...

Страница 2: ...DIRECT SUN ACTION AND FROM THE FROST MAKE SURE THAT ACCESS TO THE APPLIANCE IS ALLOWED ONLY TO QUALIFIED PERSONAL BEFORE USING THE USER HAS TO READ THIS MANUAL CAREFULLY AND TAKE NOTE OF THE WARNINGS...

Страница 3: ...INSTALLATION 9 3 2 EXCHANGE OF THE CAPILLARY 10 4 FUNCTIONING 11 4 1 FIRST START UP 11 4 2 NORMAL USE 11 4 3 MIX 11 4 4 LOW PRESSURE ALARM 11 4 5 LEAKAGE ALARM 11 4 6 INACTIVITY 11 4 7 ELECTRONIC BOA...

Страница 4: ...nly perform the operations described in this manual and they have not to perform tasks reserved to maintenance or qualified technician WARNING This equipment needs regular maintenance in order to guar...

Страница 5: ...pment in the absence of periodic maintenance IT IS STRICTLY FORBIDDEN to use the equipment in hygienic conditions or environments that aren t in compliance with those indicated in this manual IT IS ST...

Страница 6: ...able Minimum water temperature C 5 Maximum water temperature C 35 Minimum inlet capacity l h 400 Maximum inlet pressure 500 kpa Minimum inlet pressure 150 kpa LOCATION SPECIFICATIONS Maximum relative...

Страница 7: ...NNER LAYOUT 1 By pass valve and mix 2 Inlet solenoid valve 3 Maximal pressure switch 4 Minimal pressure switch 5 Permeate block 6 Membranes 7 Expansion tank 8 Expansion tank charge 9 PROFINE BLUE SMAL...

Страница 8: ...pecifications listed in the CE plate with an efficient ground and protected by 0 03 mA differential switch If the power cord is damaged it must be replaced by the manufacturer dealer or a similarly qu...

Страница 9: ...m drill bit drill a hole in the exhaust pipe of the siphon at the nut just disassembled Insert the drain pipe into the nut and tighten the nut on the bracket Tighten the nut and install the valve with...

Страница 10: ...5 BLUE 105h BLACK 95h 8 WHITE 85h 7 5 BLACK 95h 7 HARDNESS F 11 12 25 26 35 36 44 Dh 5 6 15 16 20 21 25 WARNING After replacing the capillary set the filter change time in the control unit 4 8 WARNIN...

Страница 11: ...s it with water permeate downstream of the membranes This device must be calibrated during installation Emergency BY PASS function with unplugged power supply system when it s open it lets the regular...

Страница 12: ...ssing key the software release is displayed It performs the flushing of the membranes for 2 minutes by opening the inlet solenoid valve It resets the hours of operation of the filter Confirm with It d...

Страница 13: ...the and keys and confirm with the key Select functioning by time or without time limitwith the and keys and confirm with the key Access the modification of the litres hour conversion parameter with t...

Страница 14: ...It is advisable to carry out routine maintenance at least 1 time every 6 months even if you have not reached the limit of exhaustion of the filters WARNING This equipment should be serviced periodical...

Страница 15: ...f the new cartridge in the provided space Insert the new cartridge without forcing it in any way making sure to enter the tabs in their own seats in the head and turn it 90 to the right until it stops...

Страница 16: ...smotic water or dose 250 ml of hydrogen peroxide at 24 volumes per 10 liters of osmotic water or dose 180 ml of hydrogen peroxide at 36 volumes per 10 liters of osmotic water or dose 50 ml of hydrogen...

Страница 17: ...e Reset the alarm see 4 6 If the alarm repeats or in case of water leakage close the upstream water supply valve unplug it and contact the assistance service Problem With the display turned on and wor...

Страница 18: ...in hydraulic circuit Contact assistance service Problem With dishwasher in loading phase with display on the osmosis doesn t start Cause Solution Incorrect connections The in out links were inverted M...

Страница 19: ...hazardous materials into the municipal waste stream The crossed out bin symbol reminds the user not to dispose of this product as unsorted municipal waste If disposed of privately exhausted filters a...

Страница 20: ...cribed above is in conformity with the relevant Union harmonization legislation 2006 42 UE machines 2014 30 UE EMC 2014 35 UE LVD 2011 65 UE RHOS2 References to the relevant harmonized standards used...

Страница 21: ...21 Cod 2050 0001 Rev 1 1 ATTACHMENT II MAINTENANCE LOG INSTALLATION TECHNICIAN TECHNICIAN TECHNICIAN REGULAR MAINTENANCE TECHNICIAN TECHNICIAN TECHNICIAN...

Отзывы: