background image

IMPORTANT INFORMATION: 

 File in your maintenance records

 

Maintenance Manual & Operating Instructions

CastVac

TM

CASTVAC System Control

3/06

986-1-11 Rev-L

REF 986
REF 987
REF 996

European Authorized Rep:
RA/QA Manager

Stryker France
ZAC Satolas Green Pusignan
Av. de Satolas Green
69881 MEYZIEU Cedex
France

Stryker Instruments
4100 E. Milham
Kalamazoo, Michigan
(USA) 49001
269-323-7700
1-800-253-3210
www.stryker.com

Содержание CastVac REF 986

Страница 1: ...cTM CASTVAC System Control 3 06 986 1 11 Rev L REF 986 REF 987 REF 996 European Authorized Rep RA QA Manager Stryker France ZAC Satolas Green Pusignan Av de Satolas Green 69881 MEYZIEU Cedex France Stryker Instruments 4100 E Milham Kalamazoo Michigan USA 49001 269 323 7700 1 800 253 3210 www stryker com ...

Страница 2: ...arent Take special precautions regarding electromagnetic compatibility EMC when using medical electrical equipment like the CastVac REF 996 Install and place the CastVac REF 996 into service according to the EMC information in this manual Portable and mobile RF communications equipment can affect the function of the CastVac REF 996 Use only Stryker approved accessories Other accessories may result...

Страница 3: ...tVac with 8 foot hose and Wall Mount 100 120V unit REF 996 CastVac with 8 foot hose and Mobile Stand 220 240V unit A lightening bolt within a triangle is intended to alert service personnel of the presence of dangerous voltage that may cause injury or fatal electrical shock Disconnect unit from electrical supply before servicing An exclamation point within a triangle is intended to alert the user ...

Страница 4: ... and align it with the hole in the canister connector Push hose into canister connector Release button 2 Remove the cover plug from the Cast Cutter 3 Depress the button located on the hose cuff and align with hole in Cast Cutter Gently push the vacuum hose into the hose mount Release button 4 Attach the Cast Cutter power cord to the vacuum hose straps as shown Hose Attachment 1 2 3 4 ...

Страница 5: ...ly DO NOT attach other devices The switch in the ON position enables the CastVac to run constantly The Cast Cutter operates independently and can be turned on and off while the vacuum runs steadily Models REF 986 987 996 With the switch in the STANDBY position both the Cast Cutter and CastVac are activated by the Cast Cutter switch The vacuum runs only when the Cast Cutter is switched on and shuts...

Страница 6: ...r CAUTIONS Allow vacuum to dry completely before using DO NOT IMMERSE any portion of the vacuum unit 1 Unplug vacuum unit before cleaning 2 Use a sponge dampened in soapy water or disinfectant to clean all external areas of the canister and stand DO NOT allow liquids to splash or seep inside the unit No adjustments are necessary However it is suggested that prior to use the vacuum unit should be o...

Страница 7: ...oring or transporting Troubleshooting WARNING Disconnect from electrical supply before servicing Problem Vacuum fails to start Low Suction Unit starts but stops without being turned off Action Ensure unit is plugged in and the three way switch is in the correct position Replace filter cartridge Check hose for blockage Clean filter cartridge The CastVac is provided with a thermal protective auto re...

Страница 8: ... in 620 mm Diameter 8 75 in 222 mm NEMA 5 15P Hospital Grade plug Cord Length 12 ft 3 6 m Outlet NEMA 5 15R Hospital Grade 100 120V Use with Stryker Cast Cutter models REF 840 848 940 only CLASS I IPX0 Ordinary Equipment Duty Cycle Continuous Operation CSA International CAN CSA C22 2 No 601 1 M 90 IEC 60601 1 UL 60601 1 If the equipment experiences sporadic electrical interference Turn off all ele...

Страница 9: ...IEC 61000 3 2 Voltage fluctuations flicker emissions IEC 61000 3 3 Compliance Group 1 Class A Class A Complies Electromagnetic environment guidance The CastVac REF 996 uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment The CastVac REF 996 is suitable for use in all establishments other th...

Страница 10: ...any part of the CastVac REF 996 including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance d 1 67 P d 1 67 P 80 MHz to 800 MHz d 2 33 P 800 MHz to 2 5 GHz Where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommende...

Страница 11: ...cy 50 60 Hz magnetic field IEC 61000 4 8 Compliance level 2 4 6 kV contact 2 4 8 kV air 2 kV for power supply lines 1 kV for input output lines 1 kV differential mode 2 kV common mode 95 Reduction 10ms 60 Reduction 100ms 30 Reduction 500ms 95 Reduction 5s 3 A m 50Hz CRT 1A m Electromagnetic environment guidance Floors should be wood concrete or ceramic tile If floors are covered with synthetic mat...

Страница 12: ...kHz to 80 MHz d P 80 MHz to 800 MHz d P 800 MHz to 2 5 GHz d P For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 80...

Страница 13: ... 8 ft long 940 Cast Cutter to 886 PlasterVac 940 886 USA only Stryker Instruments products are warranted to the original purchaser for a period of one year from the date of purchase to be free from defects in material and workmanship This warranty extends to all purchasers and is limited to the repair or re placement of the product without charge when returned to Stryker Instruments 4100 E Milham ...

Страница 14: ...C Satolas Green Pusignan Av de Satolas Green 69881 MEYZIEU Cedex France Stryker Instruments 4100 E Milham Kalamazoo Michigan USA 49001 269 323 7700 1 800 253 3210 www stryker com Euro Equiv 0986 001 715 N Euro Equiv 0986 001 730 Japanese Equiv 0986 001 725 ...

Отзывы: