1.  INTENDED USE  

 
The intended use of the Operatherm 202 / 200 is to prevent or minimise heat loss from patients in the 
Operating Room.   
The Operatherm must be used in compliance with the instructions in this manual and in accordance to 
accepted hospital and clinical routines

 
 

2 SAFETY 

INSTRUCTIONS 

 

Please read the complete user manual carefully before using the equipment.

 

 

W A R N I N G  

 

Never use a unit that do not perform the Self Test according to this manual.  
Never use a unit that has been dropped or fallen onto the floor before it has been examined by a 
qualified technician 
 
The Operatherm 202 is a very safe equipment. However certain combinations of heat, pressure, 
long term surgery and disinfectant fluids can cause skin damage. Therefore it is important to 
observe the advice regarding pressure relief, excessive fluid and temperature selections given 
later in this section.

 

 
  

 
 

2.2 EXPLANATION 

OF 

SYMBOLS 

 

 
 
 

Equipment not suitable for use in the presence of a flammable anaesthetic 
mixture with air or with oxygen or nitrous oxide 

 
 
 

Protection grade BF, Defibrillation safe 

 
 
 

See users manual 

~

 

AC current 

 
 

DC current 

 
 
 

ON – OFF switch. 

 
 

/ IPX1 

 

Drip proof, Control Unit 

 
 
 

/ IPX7  

 

Water tight, Heating Pad 

0413

 

Fullfills MDD 93/42 EEC.  
EC-Type examination certificate issued by a Notified Body  
( 0413= Intertech SEMKO, Sweden) 

 

Содержание Operatherm 202

Страница 1: ...NANCE 12 NOTE Patient warming equipment can harm the patient if not used correctly Never use a unit that do not perform the Self Test according to this manual Make sure that the performance of the uni...

Страница 2: ...t has been examined by a qualified technician The Operatherm 202 is a very safe equipment However certain combinations of heat pressure long term surgery and disinfectant fluids can cause skin damage...

Страница 3: ...safety 4 The Heating Pad cable may not be altered in length or repaired in any other way 5 See Users Manual section 10 concerning cleaning and maintenance of Heating Pads and Control Units 6 Sharp obj...

Страница 4: ...nt 2 8 TEMPERATURE SETTING Under normal conditions and with KanMed Gel Pads under the patient the maximum temperature 39 C should be selected During long interventions one must consider a lower temper...

Страница 5: ...Watertight except connector Surface electrical properties Not electrical conductive Control Unit Voltage 220 240 V AC 100 and110 120 V AC is also available Frequency 50 60 Hz Power 285 VA max Temperat...

Страница 6: ...is switched on it will perform a self test of all functions including the internal safety circuit This is how the self test runs When pressing the ON OFF button all front panel indicators and the beep...

Страница 7: ...ong pad or of the 4 elements in the short pad should become warmer than 40 25 C The visual and acoustic alarm will be activated and the over heated element s will be switched off for 90 sec The acoust...

Страница 8: ...ts disconnected during normal use the pad fault indicator and the 32oC LED will be lit The beeper will also sound Switch the unit off and reconnect the Heating Pad Restart the unit and observe a corre...

Страница 9: ...a technician 5 3 SELECTING TEMPERATURE In case you wish to select another temperature than 37 C then do as described in chapter 4 4 1 It will take about 8 15 minutes for the elements in the pad to re...

Страница 10: ...1 2 MAINS POWER FAILURE ALARM AND RESTARTING Connect the Heating Pad to the Control Unit and the Control Unit to the mains Switch on the unit and wait for the self test to be completed Disconnect the...

Страница 11: ...tact your local dealer 8 EMC COMPATIBILITY STATEMENT Please see www kanmed se section downloads Operatherm 9 WARRANTY STATEMENT KanMed warrants the purchaser that the Operatherm Control Unit and Heati...

Страница 12: ...ied out by the user operator Self test at Power ON Every time the equipment is switched on Visual inspection of pad cable and connector Before every use Checks carried out by technical personnel Alarm...

Отзывы: