background image

Manual KanMed Operatherm 

 

Page 

3

 

 

 

 
 
 
 

Key switch for temperatures <35ºC and >37ºC 

 
 
 

Temperature selector switch 

 
 
 

Alarm silence switch 

 
 
 

Heating Pad failure alarm 

 
 
 
 
 

Over temperature alarm 

 
 
 

Control Unit failure 

 
 
 

Earth potential equalisation connection point 

 

 

2.3  EXPLANATION OF PICTURES ON THE HEATING PAD 

 
 
 
 
 
 
 
 

Refer to users manual 

Heating Pads OP-200-xxx may  
only be used with Operatherm Control Unit OP-200/202 

Unauthorised persons are not allowed to open the Control Unit or the Heating Pad Connector. 
Any such interventions may endanger the patient safety. 

The Heating Pad cable may not be altered in length or repaired in any other way. 

See Users Manual section 10, concerning cleaning and maintenance of Heating Pads and 
Control Units 

Sharp objects must not penetrate the Heating Pad 

Place the Heating Pad flat on the operating table with the 

text side towards the patien

t. 

Gel Pads must be placed over the Heating Pad.  

When used with children, ensure that the gel pad Gel Pads covers the whole Heating Pad 

10 

Ensure that the text side of the Heating Pad is towards the patient or ECG disturbance may 
occur. 

11 

The Heating Pad should be stored using the Pad Hanger or rolled with a diameter not smaller 
than 150mm. Don't bend or fold the Heating Pad so that sharp crease arises. 

Содержание Operatherm 202

Страница 1: ...NANCE 12 NOTE Patient warming equipment can harm the patient if not used correctly Never use a unit that do not perform the Self Test according to this manual Make sure that the performance of the uni...

Страница 2: ...t has been examined by a qualified technician The Operatherm 202 is a very safe equipment However certain combinations of heat pressure long term surgery and disinfectant fluids can cause skin damage...

Страница 3: ...safety 4 The Heating Pad cable may not be altered in length or repaired in any other way 5 See Users Manual section 10 concerning cleaning and maintenance of Heating Pads and Control Units 6 Sharp obj...

Страница 4: ...nt 2 8 TEMPERATURE SETTING Under normal conditions and with KanMed Gel Pads under the patient the maximum temperature 39 C should be selected During long interventions one must consider a lower temper...

Страница 5: ...Watertight except connector Surface electrical properties Not electrical conductive Control Unit Voltage 220 240 V AC 100 and110 120 V AC is also available Frequency 50 60 Hz Power 285 VA max Temperat...

Страница 6: ...is switched on it will perform a self test of all functions including the internal safety circuit This is how the self test runs When pressing the ON OFF button all front panel indicators and the beep...

Страница 7: ...ong pad or of the 4 elements in the short pad should become warmer than 40 25 C The visual and acoustic alarm will be activated and the over heated element s will be switched off for 90 sec The acoust...

Страница 8: ...ts disconnected during normal use the pad fault indicator and the 32oC LED will be lit The beeper will also sound Switch the unit off and reconnect the Heating Pad Restart the unit and observe a corre...

Страница 9: ...a technician 5 3 SELECTING TEMPERATURE In case you wish to select another temperature than 37 C then do as described in chapter 4 4 1 It will take about 8 15 minutes for the elements in the pad to re...

Страница 10: ...1 2 MAINS POWER FAILURE ALARM AND RESTARTING Connect the Heating Pad to the Control Unit and the Control Unit to the mains Switch on the unit and wait for the self test to be completed Disconnect the...

Страница 11: ...tact your local dealer 8 EMC COMPATIBILITY STATEMENT Please see www kanmed se section downloads Operatherm 9 WARRANTY STATEMENT KanMed warrants the purchaser that the Operatherm Control Unit and Heati...

Страница 12: ...ied out by the user operator Self test at Power ON Every time the equipment is switched on Visual inspection of pad cable and connector Before every use Checks carried out by technical personnel Alarm...