5 Cleaning and Maintenance 

 

 

Recommended controls  Executed by 

Time schedule  Additional information 

Cleaning of the bed 

User 

Hospital routines  Use a surface cleaning and 

disinfecting agent recommended 
by your internal hygienist. 

Note: KANMED Baby Warming 

Bed must under no 

circumstances be showered with 
any fluid. 

Do not clean it in disinfecting 

chambers. 

Cables and connectors 

User 

Every day 

Make sure that they are in good 

condition.  

Mattress 

User 

Hospital routines  See label on Mattress edge 

Tent 

User 

Hospital routines  See label tent 

Kanmed Babywarmer  

User 

 

See Kanmed Babywarmer user 
manual 

Control of functions and 

adjustments. 

Technician 

At least once a 

year 

See service section 

 
 

 

 

6 Trouble shooting 

 
 

Symptom 

Possible cause 

Action 

No mains 

Check the cables. 
 

Something blocking the 

movement 

Check that nothing blocks 

The lifting column does not move 
as wanted. 

(not for BB-3) 

Electrical connection broken or 

not properly connected 

Technical service 

Control is out of order or needs 
adjustment 

Technical service 

You can not tilt the bed properly. 
It gets stuck in utmost positions. 

It does not keep the position 

Something is blocking it 

Check that nothing blocks 

it 

Bed feels unstable 

Loose screws or wheels 

Technical service 

Height adjustments makes 
unusual noises 

 Technical 

service 

 

7 Warranty Statement 

 

KANMED warrants the purchaser that the Babybed is free from defects in materials and workmanship for a period of 
12 month from the date of delivery. 
The sole obligation of KANMED with respect to any such defect is limited to the repair with new or re-manufactured 
parts or, at the discretion of KANMED, replacement of the equipment or refunding of the purchase price. 
This warranty shall not apply if the product has been modified, adjusted or repaired other than by KANMED or by 
organisations authorised by KANMED or modified, adjusted or repaired not in accordance with written instructions 
provided by KANMED. Neither shall the warranty apply if the equipment has been subject to misuse, negligence or 
accident.  
These warranties are made on the condition that prompt notification of a defect is given to KANMED or its authorised 
dealers within the warranty period. 
KANMED shall have the sole right to determine whether a defect exists. 
KANMED shall not in any case be liable for special or consequential damages arising from the breach of warranty, 
reach of contract, negligence or any other legal theory. 

Содержание B-1

Страница 1: ...User Manual Art no BB1 070 1 March 2007 Manufactured by KANMED AB www kanmed se Distributed by Caution Please read this manual carefully Incorrect use may cause patient injury ...

Страница 2: ...ls 3 3 Preparations before use 4 4 Handling and daily use 5 5 Cleaning and Maintenance 6 6 Trouble shooting 6 7 Warranty Statement 6 8 Technical Data 7 9 Technical service and maintenance 8 10 Spare parts Accessories 9 11 EMC guidelines 9 KANMED BABYBED TWIN VERSION AND FIXED HEIGHT VERSION ...

Страница 3: ... weight limits Do not place electronic equipment on the shelf under the bed 2 Description of Symbols Maximum Load permitted Bed surface for BB 1 and BB 3 10 kg Excl BABYWARMER Control Unit 3kg and Water mattress 4 5kg Bed surface BB 4 10 kg Excl Two BABYWARMER Control Unit 2x3kg and two Water mattresses 2x4 5kg or one BABYWARMER Control Unit 3kg and twin water mattress 9kg Shelf under bed surface ...

Страница 4: ...ent has a channel inside Thread the Tent pole into the channel in order to secure it Place the mattress in the bed 3 4 Check all bolts and screws Make sure that all bolts and screws are properly tightened and that the bed feels stable Run the electrical bed up and down and ensure that it moves smoothly without any unusual noise 3 5 Kanmed Babywarmer The Kanmed Babywarmer control unit is hung on it...

Страница 5: ... for BB 3 The height is adjustable by the pushbuttons on the front The lifting column has an overload protection that will not be activated under normal conditions Constantly adjusting height with heavy loads may activate the overload protection When the motor has cooled down the lifting column works as normal again Warning Do not block the movements when activating the electrical height adjustmen...

Страница 6: ...s not keep the position Something is blocking it Check that nothing blocks it Bed feels unstable Loose screws or wheels Technical service Height adjustments makes unusual noises Technical service 7 Warranty Statement KANMED warrants the purchaser that the Babybed is free from defects in materials and workmanship for a period of 12 month from the date of delivery The sole obligation of KANMED with ...

Страница 7: ...s I Class I NA Used standards SS EN 60 601 SS EN 60 601 Products carries the CE mark Conforms to UL 60601 1 Certified to CAN CSA C22 1 No 601 1 EMC The Kanmed Babybed contains no electronics sensitive to EMC Highly sensitive equipment connected to the same electrical outlet as the Kanmed Babybed may be affected by the lifting column electrical motor The motor may generate a moderate amount of flic...

Страница 8: ...pring from the pillar Start by taking away the small safety spring Make someone or something hold the bed Pull out the gas spring from the pillar Un tighten the locking nut by turning the gas spring you adjust the contact with the release arm 9 3 Changing side walls The easiest way to replace a side wall is to place it in its hanging down position insert one pin in the lower grove bend the wall an...

Страница 9: ... 3 Complies The Babybed is suitable for use in all establishments including domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes Guidance and manufacturer s declaration electromagnetic immunity The Babybed is intended for use in the electromagnetic environment specified below The customer or the user ...

Страница 10: ...ltage prior to application of the test level Guidance and manufacturer s declaration electromagnetic immunity The Babybed is intended for use in the electromagnetic environment specified below The customer or the user of the Babybed should assure that it is used in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Portable and mobile RF co...

Страница 11: ...ed The Babybed is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the Babybed can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the Babybed as recommended below according to the maximum output power of the communicati...

Страница 12: ...KANMED AB Gårdsfogdevägen 18B SE 168 66 BROMMA Stockholm Sweden Telephone 46 8 56 48 06 30 Telefax 46 8 56 48 06 39 E Mail info kanmed se Home page www kanmed se Distributed by ...

Отзывы: