TROUBLESHOOTING
If the display fails to stabilise at around zero within a few minutes this may
indicate an expired sensor or a sensor that has been exposed to excessive
levels of CO. If the symptom persists seek advice from Kane.
Always check for correct operation before each use. Use a known source of
CO taking care to act in a safe manner and not to expose the KANE77 sensor
to extreme temperatures or CO concentration. If the unit does not change as
expected, return the unit to Kane.
If during use, you detect CO leaks, be sure you know the appropriate actions to
take to protect yourself and others. Before switching off the monitor always
expose it to fresh air until the reading falls below 5 ppm. Note that the readings
fall much more slowly than they rise so you may need to wait some minutes.
SPECIFICATIONS
Range:
0-999 ppm
Resolution:
1 ppm
Accuracy:
+3 ppm 0 to 100 ppm
1
+5% reading > 101 ppm
Audible Alarm:
Preset at 30 ppm
Visual Indication:
LCD and LED
Battery:
9V alkaline
Battery Life
2
:
Up to 250 hours using Duracell Procell
Ambient Temperature:
0-40ºC (32-104ºF)
Warranty:
1 year
1
As per EN 50543 : 2011
2
Based on ‘normal’ use without backlight or alarms
No user serviceable parts.
Re-calibrate every 12 months.
Check for correct operation as described in this manual before every use.