TROUBLESHOOTING

If the display fails to stabilise at around zero within a few minutes this may
indicate an expired sensor or a sensor that has been exposed to excessive
levels of CO. If the symptom persists seek advice from Kane.

Always check for correct operation before each use. Use a known source of
CO taking care to act in a safe manner and not to expose the KANE77 sensor
to extreme temperatures or CO concentration. If the unit does not change as
expected, return the unit to Kane.

If during use, you detect CO leaks, be sure you know the appropriate actions to
take to protect yourself and others. Before switching off the monitor always
expose it to fresh air until the reading falls below 5 ppm. Note that the readings
fall much more slowly than they rise so you may need to wait some minutes.

SPECIFICATIONS

Range:

0-999 ppm

Resolution:

1 ppm

Accuracy:

+3 ppm 0 to 100 ppm

1

+5% reading > 101 ppm

Audible Alarm:

Preset at 30 ppm

Visual Indication:

LCD and LED

Battery:

9V alkaline

Battery Life

2

:

Up to 250 hours using Duracell Procell

Ambient Temperature:

0-40ºC (32-104ºF)

Warranty:

1 year

1

As per EN 50543 : 2011

2

Based on ‘normal’ use without backlight or alarms

No user serviceable parts.

Re-calibrate every 12 months.

Check for correct operation as described in this manual before every use.

Содержание KANE77

Страница 1: ...KANE77 Carbon Monoxide Monitor and Personal CO Alarm Stock No April 2015 Kane International Ltd ...

Страница 2: ...NG WARNING DANGER Carbon monoxide is life threatening even at relatively low concentrations Be sure you understand the risks before investigating potential leaks This monitor measures combustion gases that may be toxic in relatively low concentrations This monitor must only be used by trained and competent persons after due consideration of all the potential hazards Users of portable gas monitors ...

Страница 3: ...t automatically zeros in 50 seconds During countdown the display shows time remaining as two digits Be sure to switch on in fresh air well away from boiler flues vehicle exhausts smokers and any other potential sources of CO Turn ON Countdown The countdown starts at 50 When the countdown reaches 00 the meter will set its sensor to zero and display CO concentration using three digits A battery leve...

Страница 4: ... the CO reading at the time the HOLD button is pressed Press again to display current value CHECKING FOR CO Note Ensure that the sensor grill at the front of the monitor is free from obstruction The KANE77 can be used to check for spillage around fires or cookers but does not replicate the functions of Appliance Sampling Probes as used in the Room CO Test procedure that forms part of CPA1 The KANE...

Страница 5: ...le concentration for an 8 hour period without any known health risks according to the Department of Health and BS7967 30 ppm The maximum allowable concentration for competent persons to work in when investigating reports of fumes according to BS7967 200 ppm Mild headache fatigue dizziness and nausea after 2 to 3 hours 400 ppm Frontal headache and nausea within 1 to 2 hours Life threatening after 3...

Страница 6: ...though the reading on the display will not be the current reading but the Held or Max values BATTERY INDICATION Full battery indication Lower battery indication Do not use the monitor when the battery symbol is showing low empty Replace the battery immediately If the battery level falls to a level where the accuracy of the readings might be compromised the monitor will eventually give both audible...

Страница 7: ... at the pivot end against the spring by a few millimetres and then rotate the clip clear of the battery compartment Remove the battery compartment cover and replace the battery Alkaline Duracells are recommended Always recycle the spent battery through an approved battery recycling scheme Having fitted a fresh battery replace the cover and rotate the belt clip back to its normal position Always ch...

Страница 8: ...e appropriate actions to take to protect yourself and others Before switching off the monitor always expose it to fresh air until the reading falls below 5 ppm Note that the readings fall much more slowly than they rise so you may need to wait some minutes SPECIFICATIONS Range 0 999 ppm Resolution 1 ppm Accuracy 3 ppm 0 to 100 ppm1 5 reading 101 ppm Audible Alarm Preset at 30 ppm Visual Indication...

Страница 9: ...typically more than five years the monitor should be re calibrated and serviced annually to stop any long term sensor or electronics drift or accidental damage Local regulations may require more frequent re calibration In the UK Kane International has service facilities at Atherton near Manchester Tel 01942 873434 the primary service centre for UK customers and at Welwyn Garden City in Hertfordshi...

Страница 10: ...e the monitor has been securely packed then your package is ready for shipment back to Kane If you do not have an account with a courier company you can take your package to your local Post Office It is advisable to send the package by Special Delivery so that it is insured and traceable while in transit When we receive your monitor On receipt of your package our Service Engineers will inspect the...

Страница 11: ...artment Kane International Ltd Gibfield Park Avenue Atherton Manchester M46 0SY Northern Service Department Kane International Ltd Gibfield Park Avenue Atherton Manchester M46 0SY Northern Service Department Kane International Ltd Gibfield Park Avenue Atherton Manchester M46 0SY ...

Страница 12: ...hire AL7 1JG Your Details Name Job Title Company Name Company Address 1 Address 2 Town City County Postcode Country Phone Number Fax Number Mobile Number Email Address Product Details Note Proof of Purchase may be required for warranty claims Date Purchased as numbers 05 01 10 Purchased From Model Number KANE77 Product Serial Number ...

Страница 13: ...out Kane Magazine Advert Trade Counter Literature Personal Recommendation Internet Exhibition Other Which do you read most often Often Sometimes Hardly Ever Registered Gas Engineer Gas Installer P H P I P H A M News Heating Ventilating Plumbing Heating Plumbing Monthly Thank you for completing this survey All the information we have collected is confidential We do not sell or share data with any o...

Страница 14: ...of disturbance is acceptable If not acceptable adjust the monitor s position to minimize interference or switch off if possible the offending equipment during your test At the time of writing this manual February 2012 Kane International Ltd are not aware of any field based situation where such interference has occurred and this advice is only given to satisfy the requirements of the Directive This...

Страница 15: ...Thank you for buying this monitor Before use please register on our website www kane co uk or complete detach and return the Product Registration page ...

Отзывы: