10 KANE 358 MANUAL
GENERAL SAFETY
SAFETY WARNING
This analyser extracts combustion gases that may be toxic in relativity
low concentrations. These gases are exhausted from the bottom of the
analyser. This analyser must only be used in well-ventilated locations by
trained and competent persons after due consideration of all the
potential hazards.
Portable gas detectors users should conduct “bump” tests before
relying on units to verify atmospheres are free from hazard.
A “bump” test is a way to check an instrument works within
acceptable limits by briefly exposing it to known gas mixtures to
change the output of all sensors present.
Note: This is different from a calibration where the instrument is also
exposed to known gas mixtures but allowed to settle to a steady figure
with readings adjusted to the stated gas concentration of the test gas.
Protection Against Electric Shock (In accordance with EN 61010-1:
2010):
This analyser is designed as Class III equipment and should only be
connected to SELV circuits. The battery charger is designated as:
● Class II equipment
● Installation category II
● Pollution degree 2
● Indoor use only
● Altitude to 2000m
● Ambient temperature 0°C-40°C
● Maximum relative humidity 80% for temperatures up to 31°C
decreasing linearly to 50%RH at 40°C
● Mains supply fluctuations not to exceed 10% of the nominal voltage