flowIQ® 3100 TECHNICAL DESCRIPTION
Kamstrup
A/S • Technical Description • 5512
-1242_N1
_GB •
10.2019
57
18.3
Lifetime
•
16 years, with a single AA battery
•
Battery change is possible
18.4
Connection of Pulse Adapter
From the factory the meter is configured not to transmit optical pulses
–
otherwise, battery life will be shortened.
Therefore, having mounted the Pulse Adapter, the installer must set up the meter to
‘pulse adapter mode’
for
transmitting optical pulses.
The procedure is as follows:
1.
Unfasten the screw of the adapter ring to allow mounting of the unit on the meter
2.
Dismount the two screws at the top and remove the cover
3.
Push the button on the PCB, which is placed under the cover (see figure below).
As soon as the button is pushed, serial optical communication between Pulse Adapter and meter starts.