Kamoer Fluid Tech(Shanghai) Co.,Ltd Reef Master CMK User Manual
38
b.
Click to clear all pipelines. Here, the liquid in the built-in pipelines and colorimetric dishes on
the device is cleared. The waste liquid generated by clearing the pipelines is discharged from
the waste liquid pipeline
Test Value Correct
During the use of the equipment, if it is suspected that the detected values are inaccurate,
calibration can be carried out. The principle of calibration is to use a standard solution to calibrate
the detection results of the equipment. When the detection value of calcium and magnesium KH is
not accurate, the detection value can be corrected. The corrected reference value can be the value
of the test standard solution or the value of other considered accurate testing equipment. It is
recommended to use the standard solution for calibration.
When testing, prepare the standard solution, such as using a calcium standard solution. First,
connect the standard solution to the sample tube of the device, and then run the test through the
App. One test is not enough, and it needs to be run several more times. After the data is stable,
enter the value of the standard solution into the App in System Settings ->Maintenance ->Test Value
Correct to complete the calibration.
The calibration process is as follows:
1.
Use standard solution instead of seawater for Reef Master testing
2.
After testing several stable values, enter the App and input the value of the standard
solution into the App
Содержание Reef Master CMK
Страница 1: ...Reef Master CMK User Manual A1 2023 11 27 Kamoer Fluid Tech Shanghai Co Ltd www kamoer com...
Страница 28: ...Kamoer Fluid Tech Shanghai Co Ltd Reef Master CMK User Manual 27 Reef Master linked calcium reactor pump...
Страница 42: ...Kamoer Fluid Tech Shanghai Co Ltd Reef Master CMK User Manual 41 Equipment Maintenance...
Страница 45: ...Kamoer Fluid Tech Shanghai Co Ltd Reef Master CMK User Manual 44 Appendix...