34

 

 

 

9.

 SERVICE AND 

 

    MAINTENANCE 

 

 
 
 

 

WARNING

 

 

Before any service or cleaning be sure to disconnect the power supply 

                cable from the power supply!

 

 

Maintenance and service within the warranty period may only be performed by the authorized service 

person. 
Please advise Kambic in before alternation & modification. It is not allowed to change or rebuild any 

part of the construction without producer's written permit. 
All spare parts are in stock in the producer's magazine. It is only allowed to use the original spare 

parts. 

Maintenance  and  service  within  the  warranty  period  may  only  be  done  by  the  producer’s  service 

department or authorized service. 
With proper care the instrument should require very little maintenance. Keep the interior and the 

exterior clean. For cleaning use some soft cleaning detergents and soft cloth. 

Before starting the process check the level of distil water. Both level switches must be 

covered with water! Use only distilled water! 

Regularly check parts of cooling system and system for maintaining relative humidity. 

 
At least once per six month check and clean: 

 

-

 

Condenser blades (for cleaning use compressed air. (Blow from behind) 

-

 

Release distilled water from water reservoir and if necessary clean reservoir  

-

 

Release distilled water from water pool in chamber and clean pool 

 
 
 
 
 
 
 
 
 
 
 

Water reservoir 

Drain valve from reservoir 

 

Cooling system condenser 

(to be cleaned with compressed air) 

 

In the event of failure, call the service;  

Please find contacts on page 1 or on our web site: 

www.kambicmetrology.com

  

Содержание KK-105 CH

Страница 1: ...User Manual Climatic Chamber KK 105 CH...

Страница 2: ...t profits loss of income loss of business opportunities loss of use and other related exposures caused by incorrect use of product Warranty is valid for 13 months from date of shipping and includes sp...

Страница 3: ...SETTINGS AUTOMATIC 22 7 SPECIAL SETUP 26 7 1 TIME AND DATE 26 7 2 CORRECTION SETUP 27 7 3 CORRECTION SETUP SCHEME 28 7 4 USERS PASSWORDS 31 7 5 PROGRAM COPY 32 8 CONTROL PANEL DESCRIPTION 33 9 SERVIC...

Страница 4: ...t sure about safe use of the product in your specific case do not hesitate to contact your dealer or us via info kambic com for further information or opinions In no event shall Kambic d o o be held l...

Страница 5: ...operate the equipment in wet and condensing environment STABILITY HAZARD Do not load or affect product stability with heavy loads or climbing the product TRANSPORT HAZARD Transport the chamber only in...

Страница 6: ...ere Verwendung des Ger tes in Ihrem spezifischen Fall unsicher sind z gern Sie nicht Ihren H ndler oder uns ber info kambic com zu weiteren Informationen oder Stellungnahmen zu kontaktieren Keinesfall...

Страница 7: ...ensierender Umgebung betreiben GEFAHR DER SCHWERPUNKTVERLAGERUNG Keine schweren Lasten laden oder Stabilit t durch Zuladung oder Besteigen des Produktes beeinflussen GEFAHR DURCH TRANSPORT Transportie...

Страница 8: ...z des doutes concernant une utilisation s re dans votre cas pr cis n h sitez pas contacter votre vendeur ou nous m mes via info kambic com pour plus d informations ou de conseils En aucun cas Kambic d...

Страница 9: ...environnement humide et pr sentant de la condensation RISQUE D INSTABILIT N affectez pas la stabilit du produit avec de lourdes charges ou en montant sur celui ci RISQUES LI S AU TRANSPORT Ne transpo...

Страница 10: ...est clara o tiene dudas en cuanto al uso seguro del producto en su caso particular contacte con su distribuidor o escr banos a info kambic com para obtener informaci n adicional u opini n Sociedad li...

Страница 11: ...No manipule el equipo en un ambiente h medo y de condensaci n RIESGO DE INESTABILIDAD No aplique carga ni desestabilice el producto con cargas pesadas No subir al producto RIESGOS AL TRANSPORTAR Tran...

Страница 12: ...User Manual KK 105 CH 12 info kambic com Kambic d o o o...

Страница 13: ...User Manual KK 105 CH 13...

Страница 14: ...he chamber and the door is made of high quality silicon Controller Climatic chamber is equipped with microprocessor controller which is specially designed for simulation of climatic conditions Control...

Страница 15: ...ow pipe Record process Appliance has a built in RS232 port that is used to communicate with PC or printer The printer used when we require documented report of the process The record is performed by a...

Страница 16: ...ure range 5 C to 180 C Temperature stability 0 3 C Humidity range 10 98 Rh Humidity stability 3 Rh Control microprocessor controller DPC 420 Cooling Air cooled compressor aggregate Over temperature sh...

Страница 17: ...your action requires longer period of open doors switch off the unite for this period Microprocessor controller has a special safety system which turns heaters off if temperature rises for more than...

Страница 18: ...has to be replaced it must be manually placed on a transport palette and only then lifted and transported with a vehicle Installation must be done in a dry well ventilated room Be sure that the seal...

Страница 19: ...he time count down starts See special settings Tc 38 3 C 40 C Current chamber temperature is 38 3 C set chamber temp 40 C Rh 62 2 75 Current chamber humidity is 62 2 set humidity is 75 Tw 29 8 C 34 7...

Страница 20: ...ual program can be set while working 1 START PROGRAM 2 MANUAL 3 PROGRAM SETUP 4 SPECIAL SETUP MENU 1 SELECT PROGRAM 0 MANUAL 1 50 AVTO Program name PROGRAM 0 MANUAL PROG 0 SETUP MANUAL CHAMBER TEMPERA...

Страница 21: ...red relative humidity in a chamber On display you can see an available humidity range in a round bracket valid for water dew point system 5 Fan speed You can set the desired chamber air circulation fa...

Страница 22: ...ame PROGRAM 1 AVTO PROG 1 SETUP RECORDING PERIOD 3 min PROG 1 SETUP DELAYED AUTOMATIC START WITHOUT DELAY PROG 1 SETUP DATE OF DELAYED START OFF PROG 1 SETUP STEP 1 CHAMBER TEMPERATURE 40 C PROG 1 SET...

Страница 23: ...steps that need to be executed 7 current step in process Example number of steps is set to 5 cycling of steps is set to 3 cycle from step is 5 cycle to step is set to 1 So the program shall do 5 step...

Страница 24: ...change value press ATTENTION The hour of the delayed start should always be set in combination with date of delayed start ATTENTION The date of the delayed start should always be set in combination w...

Страница 25: ...duration time for step 1 in time range from 1 min to 1440 min Duration time can also be set endless under 1 min or above 1440 min If you selected more than 1 step then parameters for other steps must...

Страница 26: ...menu Display Press number to enter sub menu time setup Display To change value under blinking line press to move blinking line press To save and exit setting press 1 START PROGRAM 2 MANUAL 3 PROGRAM...

Страница 27: ...finish with enter 1 START PROGRAM 2 MANUAL 3 PROGRAM SETUP 4 SPECIAL SETUP 4 1 TIME SETUP 2 CORRECTION SETUP 3 USERS PASSWORDS 4 PROGRAM COPY 2 MENU ENTER PASSWORD AND FINISH WITH ENTER PASSWORD 0 0 A...

Страница 28: ...UNDER 35 C SPECIAL SETUP PRESSURE IN COOLING SYSTEM 9 5 bar SPECIAL SETUP ALARM TOLERANCE TEMPERAT ALARM OFF Pheat chamb 0 0 Pcool chamb 0 0 Pheat water 0 0 Pcool water 0 0 SPECIAL SETUP CICLE SERIAL...

Страница 29: ...erature step The countdown will start when temperature is in set limits For example Set temperature 50 C Set time 60 min Set START TIME COUNT IN RANGE 1 C The countdown will start when the temperature...

Страница 30: ...avoid water condensation on the doors a weak heater is installed on inner side of doors and in the glass In case of inconvenient condensation at some temperatures you may higher lower this value 24 Pr...

Страница 31: ...o technicians Technician password allows us to access all operator functions settings of programs and special setup settings The users passwords settings are not accessible and they are not allowed to...

Страница 32: ...this menu press button Display Press to enter special setup menu Display Press to enter program copy setup menu Display DATA FROM PROG 1 Data from selected program number 1 will be copied COPY TO PRO...

Страница 33: ...User Manual KK 105 CH 33 8 CONTROL PANEL DESCRIPTION...

Страница 34: ...h proper care the instrument should require very little maintenance Keep the interior and the exterior clean For cleaning use some soft cleaning detergents and soft cloth Before starting the process c...

Страница 35: ...wall It must also be in a dry and well ventilated room Additional space or relocation of the apparatus may be necessary when carrying out any maintenance or servicing The user must provide electrical...

Страница 36: ...ste Instead you should protect human health and the environment by handing over your waste equipment to a designated collection point for the recycling of waste electrical and electronic equipment For...

Страница 37: ...of an error on this system a mechanical over temperature shut off device will disable heating If this happens the cause of overheating must be found and the over temperature shut off device must be ma...

Страница 38: ...User Manual KK 105 CH 38 13 ATTACHMENTS 13 1 Wiring diagram 13 2 RS 232 Communication protocol 13 3 Declarations...

Отзывы:

Похожие инструкции для KK-105 CH