1-5: Temperature probe correction
•
For temperature probe 5 corrections points are defined.
For each correction point correction can be set. To change value use .
6-10: Water temperature probe correction
•
For water temperature probe also 5 corrections points are defined.
For each correction point correction can be set. To change value use .
11: Time count-down tolerance (temperature)
•
You can set the time countdown tolerance of each temperature step. The
countdown will start when temperature is in set limits.
For example:
Set temperature: 50°C
Set time: 60 min
Set START TIME COUNT IN RANGE : +-
1°C
The countdown will start when the temperature will
be between 49.0°C and 51.0°C !
12: Time count-down tolerance (humidity)
•
You can set the time countdown tolerance of each humidity step. The countdown
will start when humidity is in the set limits.
For example:
Set humidity: 50%
Set time: 60 min
Set COUNT DOWN TOLERANCE IN HUMIDITY RANGE: +-5%
The countdown will start when humidity will be between 45% and 55% !
ATTENTION:
Time count-down begins when both parameters (temperature and humidity) are in their limits!