WSMANUAL 0114

File>>Open Cal File Window

 – Allows users to 

work on saved calibration file or open up a new one. 

If you have not selected a communication port, you 

will not be allowed to upload or download a calibra-

tion. See Fig. 4-5.

File>>Unit Software Reset – 

Resets the Weight Scale 

Controller.

 

File>>Page Setup 

 – Allows users configure page 

setup for printer output

File>>Print

 – Prints contents of main window

Comm>> Open

  – Opens data window. See Page 9, 

Step 8.

Comm>>Close

 – Closes data window

Comm>>Parameters 

– Allows user to select and 

configure serial port. See Page 9, Steps 5-7.

Comm>>Terminal Window 

 – Allows user to see 

all incoming messages from Weight Scale Controller. 

Also allows user to send text commands to the WSC 

from the ‘Send’ window. A send history is also pro-

vided to ease repetitive tasks. Note: This functional-

ity requires selection of the serial port. See 4-6.

O P E R AT I O N   c o n t i n u e d

CONTROLLER SOFTWARE MENUS

File>>Enter Password 

- A pop-up window will appear 

for password entry. Once the password has been cor-

rectly entered, additional data and inputs will appear in 

the main screen. See Fig.  4.4.

FIG. 4-4

FIG. 4-5

FIG. 4-6

Содержание Simple Precision Weight Scale

Страница 1: ...IANT TEL 1 713 784 0000 FAX 1 713 784 0001 Email Sales Kam com KAM CONTROLS INC 3939 Ann Arbor Drive Houston Texas 77063 USA www KAM com User Manual Kam Weight Scale WSMANUAL 0114 An ISO 9001 certifie...

Страница 2: ...ion 2 Available Models 2 Additional Specifications 2 2 Dimensional Drawings 3 5 3 Installation 6 Placement 6 7 Wiring 7 8 4 Operations 9 Connecting the RS232 9 Controller Software Menus 10 11 KAM CONT...

Страница 3: ...ale unit communicates with the KAM SC Sampler Controller via 4 20 mA The 4 20 mA signal directly correlates to the current weight of the Sample Receiver and its volumetric equivalent In this way op er...

Страница 4: ...02 lb 9 grams Dimensions 9 25 W x 9 25 D x 1 25 H 235 MM x 235 MM x 32 MM Maximum weight Recommended no more than 66 lbs 30 KG max overload 90 lbs 41 KG 0 100 lbs 0 27 KG Standard for use with a 3 or...

Страница 5: ...odel MODELS AND SPECIFICATIONS Continued 0 300 lbs 0 27 KG Standard for use with a 3 or 5 gallon sample receiver Resolution 1 lb 45 grams resolution Dimensions 14 W x 14 D x 1 25 H 406 MM x 305M M x 3...

Страница 6: ...Stainless steel and Aluminum Supply Voltage 10 24 VDC Signal Output 4 20 mA Temperature Range 0 185 F 18 85 C Sensitivity 2 mV Volts 10 Controller Dimensions 5 1 W x 4 5 D x 6 6 H 130MM x 114MM x 168...

Страница 7: ...ess Pipeline Flow Weight Scale KAM SC Sample Controller Power Source Field Control Room Sample Receiver 1 Sample Receiver 2 Flowmeter KAM IAS Isokinetic Automatic Sampler Sample Discharge Heater 3 way...

Страница 8: ...NING The KAM Weight Scale has mechanical stops to prevent damage from overweight receivers HOWEVER placing receivers above the maximum weight or slamming receivers down onto scale will result in damag...

Страница 9: ...onnect the negative wire to the V terminal 6 Connect 4 20mA output wires to Loop and Loop 7 Reconnect LCD cable going to display plate and secure with 3 4 40 screws Fig 3 5 8 Power can now be turned o...

Страница 10: ...ou will nee a computer with a serial port or serial port converter and installed WSC software NOTE In order to change settings in the control ler user must first select the communications port While s...

Страница 11: ...oses data window Comm Parameters Allows user to select and configure serial port See Page 9 Steps 5 7 Comm Terminal Window Allows user to see all incoming messages from Weight Scale Controller Also al...

Страница 12: ...nsume disk space Reporting Select Logging Directory Allows user to select a directory to save the log files creat ed by the program The default is C ProgramData Kam WSC LOGS See Fig 4 7 Reporting Gene...