19

KFMANUAL 0317

KAM CONTROLS, INC.

MAINTENANCE continued

The unit can be stored for up to 1 week without use without needing to remove the reagents 

or disassemble. Do not disconnect any of the attachments on the Generator Solution Cell and 

keep all parts mounted to prevent moisture absorption in the reagents. If the cell is left stand-

ing with the drying tube removed, the ceramic frit on the Cathode Solution Cell will absorb 

moisture.

For storage/no usage for over 1 week, the unit needs to be disassembled and all glass needs 

to be cleaned  with soap and hot water,  rinsed, and dried.  Leaving the reagents in for an 

extended period of time can lead to crystallization. 

To keep the Generator Solution Cell performing accurately and effectively be sure that a thin 

layer of Special Grease is applied to all glass tapered joints each time the Reagents are 

replaced. Frequent greasing will prevent the glassware from seizing and causing damage to 

the cell. The Special Grease does not absorb moisture, so it will not hinder the results of the 

analysis. 

NOTE: No other grease should be used with a KAM

®

 KF Karl Fischer Moisture Analyzer due 

to the potential for the other types of grease to absorb moisture and interfere with titration if 

it is silicon based. KAM CONTROLS is not responsible for problems resulting from the use of 

improper grease. 

STORAGE

SPECIAL GREASE

CATHODE CELL ELECTRODE CLEANING PROCEDURES

Dirty electrodes (platinum grids) usually result in the following problems:

• A decrease in titration efficiency resulting in a longer measuring time

• An increase in the blank (residual) value of titration current due to absorption of moisture

   in  the dirty area

• Unstable titration current and lack of reproducible results

• Staining on the ceramic frit at the bottom of the Cathode Solution Cell

NOTE: KAM

®

 recommends the use of chromic sulfuric acid (3 ml of potassium dichro-

mate+400 ml of reagent grade sulfuric acid). Because nitric acid produces corrosive fumes, be 

sure and conduct all parts of this procedure in an appropriately ventilated environment.

Acid should NOT be reused. Follow appropriate disposal procedures.

KAM

®

 performs this service as a part of annual unit recertification and highly recommends end 

users send unit to factory for electrode cleaning if a suitable environment and/or chemicals are 

not available. 

Follow steps for glassware disassembly on page 18 of this manual.

Содержание Simple Precision LKF

Страница 1: ...PKF Karl Fischer Moisture Analyzer TEL 1 713 784 0000 FAX 1 713 784 0001 Email Sales KAM com KAM CONTROLS INC 3939 Ann Arbor Drive Houston Texas 77063 USA User Manual KFMANUAL 0317 An ISO 9001 certif...

Страница 2: ...dding reagent 7 Base assembly 8 Setting date and time 9 Reset reagent life 9 Sample size and titration mode 10 Initial titration 10 Running samples 11 moisture content determination Sample dilution 12...

Страница 3: ...e maintenance program By analyzing lube oil on compressors turbines etc the Moisture Analyzer can detect cooling water leaks before they become severe enough to damage equipment Karl Fischer titrimetr...

Страница 4: ...exceeds API MPMS 10 9 ASTM D4928 ISO 12937 and EI 386 Generator electrode config With diaphragm Titration speed 1000 g H2 O min max at high H2 O concentrations Power requirements Operates on either A...

Страница 5: ...ion Cell Assembly Titration Cell Sleeve Power Cord Storage Silica Gel Storage Detector Electrode Wire Cathode Solution Cell Wire Swivel Clamp 3 COMPONENTS FRONT PANEL portable model shown REAR PANEL p...

Страница 6: ...screen Battery Life Indicator blinks when charging USB Indicator BlueTooth Power Cord Charging portable units LCD Display main menu shown Screen Navigation Arrows ENTER Key Karl Fischer Moisture Anal...

Страница 7: ...o not apply pressure to the Electrode Rather let it seat under its own weight Rotate it half a turn backward and forward to ensure the grease is spread evenly and a seal has been created Drop the cera...

Страница 8: ...n Cell Jack black Detector Electrode Jack red Turn the swivel clamp toward back of KF unit and place the entire Generator Solution Cell assembly into the Titration Cell Sleeve Turn Swivel Clamp over t...

Страница 9: ...set time navigate to time indicator and use the UP DOWN arrows to change the hour 24 hour clock and use the RIGHT LEFT arrows to move between fields Press ENTER to save values and return to Main Menu...

Страница 10: ...5 After long periods without use the unit has been turned off and or each time the reagents have been replaced you must run an initial titration to remove any ambient moisture from the reagents and c...

Страница 11: ...m the inside of the syringe The tip of the plunger should exactly align with the sample volume marking Press NEW SAMPLE key on Keypad The display will read Inject Sample and a thirty second countdown...

Страница 12: ...h sample log Data for any individual sample can be retrieved by using the UP DOWN arrows to navigate to the desired sample number and pressing ENTER The display will show the date time water percentag...

Страница 13: ...ill appear Type in the desired name for the output file in the format Filename xls and hit save Example August titrations xls A window will open with all the sample files listed Select all the KFF fil...

Страница 14: ...G WITHTHE OPTIONAL BLUETOOTH PRINTER 1 2 3 4 5 MENU Reagent Life Date Time Saved Samples Language Calibration Version Info English 78 Aug15 98 BT Mode 09 21 BT Mode Modem Printer Scanning for Printer...

Страница 15: ...ueTooth icon will appear at the top of the screen This indicates the devices have been paired successfully To print from the MAIN MENU use the navigation arrows to scroll down to Saved Samples and pre...

Страница 16: ...nother record to the printer while a record is still printing NOTE if at any time the KF unit is turned off you must turn off the printer then follow the steps on page 14 of this manual to reconnect t...

Страница 17: ...on the side of the printer two times in quick succession The printer will shut down Wait until the blue light on the printer stops blinking and turns off Then press the power button once more The pri...

Страница 18: ...om the cell Dispose of fluids according to SDS sheet s Empty contents of Generator Solution Cell by pouring into a suitable container or aspiration Dispose of fluids according to the SDS sheet Thoroug...

Страница 19: ...KF Karl Fischer Moisture Analyzer due to the potential for the other types of grease to absorb moisture and interfere with titration if it is silicon based KAM CONTROLS is not responsible for problems...

Страница 20: ...night discard the nitric acid in the cell and rinse the inside and outside of the cell with running tap water until the yellow color disappears Place distilled water in the cell and place it in a 400...

Страница 21: ...Calibration Verification Check Solution 1 1 ml and 10 l Gastight Syringes 14 and 19 gauge Needles package of rubber Septums 100 Funnels 2 Special Grease 1 and Silica Gel 1 REAGENTS and CONSUMABLES GL...

Страница 22: ...SAMPLE INJECTION PORT P 09044 STIR BAR P 09050 DRYING TUBE P 09029 RUBBER SEPTUM pack of 100 P 09015 SPECIAL GREASE 25 g tube P 370420 REPLACEMENT RECHARGEABLE BATTERY for model 09031 P 370400 POWER...

Отзывы: