12

KFMANUAL 0317

KAM CONTROLS, INC.

ACCESSING SAMPLE DATA

The KAM

®

 KF saves data for up to 100 samples in 

internal memory. Samples are automatically time 

stamped and numbered in chronological order. 

Saved sample data can be retrieved at any time by 

following the steps below.

Sample79

Date: 08/15/2016

Time: 09:45

Water Cont.: 0.1098%

Mode: %

Sample Size: 0.5

Reagent age:    

     0 Days

Volume injected: 

    0.65 ml

Saved Samples

Sample74

08/14/2016

Sample73

08/14/2016

Sample72

08/14/2016

08/15/2016

Sample75

08/15/2016

Sample76

08/15/2016

Sample77

08/15/2016

Sample78

08/15/2016

Sample79

4. OPERATION continued

Return to the Main Menu by pressing the MENU 

key on the keyboard. 

Use the UP/DOWN arrows to navigate to Saved 

Samples and press the ENTER button. 

The sample log will display samples in 

chronological order. 

Use LEFT/RIGHT arrows on keyboard to page 

through sample log. 

Data for any individual sample can be retrieved 

by using the UP/DOWN arrows to navigate to the 

desired sample number and pressing ENTER. The 

display will show the date, time, water percentage, 

and sample size of the sample in question. The 

next or previous sample can be viewed using the 

UP/DOWN arrows.

Hit ENTER  to return to Sample Log/Saved 

Samples.

1.

2.

3.

4.

5.

6.

1.

2.

3.

SAMPLE DILUTION METHOD

In cases where the viscosity of the sample is such that it cannot be pulled into a syringe or 

when the anticipated water content is higher than 5%, users may dilute the sample 

according to the following steps.

From the Main Menu scroll down to Titration Mode and use RIGHT/LEFT arrows to switch to 

Dilution Mode indicated by “Dil”.

Dilute (mix) the sample with a ratio of 1 part (1 ml) sample to 99 parts (99 ml) of 

reagent-grade xylene.

The analyzer automatically adjusts for the dilution factor based on the sample size. 

MENU

Start Titration

Sample Size

Titration Mode

Reagent Life

Date & Time

Saved Samples

Language

English

78

Aug15

98%

ml

1.0

09:21

Содержание Simple Precision LKF

Страница 1: ...PKF Karl Fischer Moisture Analyzer TEL 1 713 784 0000 FAX 1 713 784 0001 Email Sales KAM com KAM CONTROLS INC 3939 Ann Arbor Drive Houston Texas 77063 USA User Manual KFMANUAL 0317 An ISO 9001 certif...

Страница 2: ...dding reagent 7 Base assembly 8 Setting date and time 9 Reset reagent life 9 Sample size and titration mode 10 Initial titration 10 Running samples 11 moisture content determination Sample dilution 12...

Страница 3: ...e maintenance program By analyzing lube oil on compressors turbines etc the Moisture Analyzer can detect cooling water leaks before they become severe enough to damage equipment Karl Fischer titrimetr...

Страница 4: ...exceeds API MPMS 10 9 ASTM D4928 ISO 12937 and EI 386 Generator electrode config With diaphragm Titration speed 1000 g H2 O min max at high H2 O concentrations Power requirements Operates on either A...

Страница 5: ...ion Cell Assembly Titration Cell Sleeve Power Cord Storage Silica Gel Storage Detector Electrode Wire Cathode Solution Cell Wire Swivel Clamp 3 COMPONENTS FRONT PANEL portable model shown REAR PANEL p...

Страница 6: ...screen Battery Life Indicator blinks when charging USB Indicator BlueTooth Power Cord Charging portable units LCD Display main menu shown Screen Navigation Arrows ENTER Key Karl Fischer Moisture Anal...

Страница 7: ...o not apply pressure to the Electrode Rather let it seat under its own weight Rotate it half a turn backward and forward to ensure the grease is spread evenly and a seal has been created Drop the cera...

Страница 8: ...n Cell Jack black Detector Electrode Jack red Turn the swivel clamp toward back of KF unit and place the entire Generator Solution Cell assembly into the Titration Cell Sleeve Turn Swivel Clamp over t...

Страница 9: ...set time navigate to time indicator and use the UP DOWN arrows to change the hour 24 hour clock and use the RIGHT LEFT arrows to move between fields Press ENTER to save values and return to Main Menu...

Страница 10: ...5 After long periods without use the unit has been turned off and or each time the reagents have been replaced you must run an initial titration to remove any ambient moisture from the reagents and c...

Страница 11: ...m the inside of the syringe The tip of the plunger should exactly align with the sample volume marking Press NEW SAMPLE key on Keypad The display will read Inject Sample and a thirty second countdown...

Страница 12: ...h sample log Data for any individual sample can be retrieved by using the UP DOWN arrows to navigate to the desired sample number and pressing ENTER The display will show the date time water percentag...

Страница 13: ...ill appear Type in the desired name for the output file in the format Filename xls and hit save Example August titrations xls A window will open with all the sample files listed Select all the KFF fil...

Страница 14: ...G WITHTHE OPTIONAL BLUETOOTH PRINTER 1 2 3 4 5 MENU Reagent Life Date Time Saved Samples Language Calibration Version Info English 78 Aug15 98 BT Mode 09 21 BT Mode Modem Printer Scanning for Printer...

Страница 15: ...ueTooth icon will appear at the top of the screen This indicates the devices have been paired successfully To print from the MAIN MENU use the navigation arrows to scroll down to Saved Samples and pre...

Страница 16: ...nother record to the printer while a record is still printing NOTE if at any time the KF unit is turned off you must turn off the printer then follow the steps on page 14 of this manual to reconnect t...

Страница 17: ...on the side of the printer two times in quick succession The printer will shut down Wait until the blue light on the printer stops blinking and turns off Then press the power button once more The pri...

Страница 18: ...om the cell Dispose of fluids according to SDS sheet s Empty contents of Generator Solution Cell by pouring into a suitable container or aspiration Dispose of fluids according to the SDS sheet Thoroug...

Страница 19: ...KF Karl Fischer Moisture Analyzer due to the potential for the other types of grease to absorb moisture and interfere with titration if it is silicon based KAM CONTROLS is not responsible for problems...

Страница 20: ...night discard the nitric acid in the cell and rinse the inside and outside of the cell with running tap water until the yellow color disappears Place distilled water in the cell and place it in a 400...

Страница 21: ...Calibration Verification Check Solution 1 1 ml and 10 l Gastight Syringes 14 and 19 gauge Needles package of rubber Septums 100 Funnels 2 Special Grease 1 and Silica Gel 1 REAGENTS and CONSUMABLES GL...

Страница 22: ...SAMPLE INJECTION PORT P 09044 STIR BAR P 09050 DRYING TUBE P 09029 RUBBER SEPTUM pack of 100 P 09015 SPECIAL GREASE 25 g tube P 370420 REPLACEMENT RECHARGEABLE BATTERY for model 09031 P 370400 POWER...

Отзывы: