ACCESSING SAMPLE DATA
The KAM
®
KF saves data for up to 100 samples in
internal memory. Samples are automatically time
stamped and numbered in chronological order.
Saved sample data can be retrieved at any time by
following the steps below.
Sample79
Date: 08/15/2016
Time: 09:45
Water Cont.: 0.1098%
Mode: %
Sample Size: 0.5
Reagent age:
0 Days
Volume injected:
0.65 ml
Saved Samples
Sample74
08/14/2016
Sample73
08/14/2016
Sample72
08/14/2016
08/15/2016
Sample75
08/15/2016
Sample76
08/15/2016
Sample77
08/15/2016
Sample78
08/15/2016
Sample79
4. OPERATION continued
Return to the Main Menu by pressing the MENU
key on the keyboard.
Use the UP/DOWN arrows to navigate to Saved
Samples and press the ENTER button.
The sample log will display samples in
chronological order.
Use LEFT/RIGHT arrows on keyboard to page
through sample log.
Data for any individual sample can be retrieved
by using the UP/DOWN arrows to navigate to the
desired sample number and pressing ENTER. The
display will show the date, time, water percentage,
and sample size of the sample in question. The
next or previous sample can be viewed using the
UP/DOWN arrows.
Hit ENTER to return to Sample Log/Saved
Samples.
1.
2.
3.
4.
5.
6.
1.
2.
3.
SAMPLE DILUTION METHOD
In cases where the viscosity of the sample is such that it cannot be pulled into a syringe or
when the anticipated water content is higher than 5%, users may dilute the sample
according to the following steps.
From the Main Menu scroll down to Titration Mode and use RIGHT/LEFT arrows to switch to
Dilution Mode indicated by “Dil”.
Dilute (mix) the sample with a ratio of 1 part (1 ml) sample to 99 parts (99 ml) of
reagent-grade xylene.
The analyzer automatically adjusts for the dilution factor based on the sample size.
MENU
Start Titration
Sample Size
Titration Mode
Reagent Life
Date & Time
Saved Samples
Language
English