RUNNING SAMPLES / MOISTURE CONTENT DETERMINATION

4. OPERATION continued

1.

2.

3.

Sample preparation and methodology must conform to the 

standards dictated in API MPMS Chapters 8.3 and 10.9. 

To draw a sample, first flush the syringe with sample 

fluid. Draw this sample into the syringe and 

discharge the sample into a waste container a 

minimum of three times. 

After the syringe has been properly flushed, draw a 

new sample in to the syringe. The sample volume 

drawn should be slightly larger than the desired 

amount.

Invert the syringe so that the needle end is point-

ing up. Wrap the needle in a paper towel (to absorb 

discharged excess fluid) and hold syringe by needle 

end.  Tap the body of the syringe to allow any air bub-

bles to rise to the top. Slowly press plunger to dis-

charge air bubbles and excess sample fluid from the 

inside of the syringe. The tip of the plunger should 

exactly align with the sample volume marking.  

Press NEW SAMPLE key on Keypad.

The display will read “Inject Sample” and a thirty-

second countdown will begin.

Insert syringe needle through the Sample Injection 

Port, piercing the Septum and continuing into the 

cell until the needle tip is fully submerged in the 

Reagent. 

Inject the sample into the Reagent. This should be 

done in a single, swift movement, but do not allow 

the sample to splash up against the wall of the 

Generator Solution Cell or the needle to touch the 

sides of the cell. 

Electrolysis begins automatically by the injection of 

the sample. When the analysis is completed, the 

unit beeps and the results will be displayed on the 

Titration Complete screen.  

09:21

(30 Sec)

Aug 15, 2016

Sample # 79 / Size: 0.5

Inject Sample

Detection
Titration

0.0000%

Sample 

Injection Port

09:21

Aug 15, 2016

Sample # 79 / Size: 0.5

Titration Complete

Detection
Titration

0.1098%

4.

5.

6.

7.

8.

09:21

Aug 15, 2016

Sample # 79 / Size: 0.5

Processing

Detection
Titration

0.0574%

Содержание Simple Precision LKF

Страница 1: ...PKF Karl Fischer Moisture Analyzer TEL 1 713 784 0000 FAX 1 713 784 0001 Email Sales KAM com KAM CONTROLS INC 3939 Ann Arbor Drive Houston Texas 77063 USA User Manual KFMANUAL 0317 An ISO 9001 certif...

Страница 2: ...dding reagent 7 Base assembly 8 Setting date and time 9 Reset reagent life 9 Sample size and titration mode 10 Initial titration 10 Running samples 11 moisture content determination Sample dilution 12...

Страница 3: ...e maintenance program By analyzing lube oil on compressors turbines etc the Moisture Analyzer can detect cooling water leaks before they become severe enough to damage equipment Karl Fischer titrimetr...

Страница 4: ...exceeds API MPMS 10 9 ASTM D4928 ISO 12937 and EI 386 Generator electrode config With diaphragm Titration speed 1000 g H2 O min max at high H2 O concentrations Power requirements Operates on either A...

Страница 5: ...ion Cell Assembly Titration Cell Sleeve Power Cord Storage Silica Gel Storage Detector Electrode Wire Cathode Solution Cell Wire Swivel Clamp 3 COMPONENTS FRONT PANEL portable model shown REAR PANEL p...

Страница 6: ...screen Battery Life Indicator blinks when charging USB Indicator BlueTooth Power Cord Charging portable units LCD Display main menu shown Screen Navigation Arrows ENTER Key Karl Fischer Moisture Anal...

Страница 7: ...o not apply pressure to the Electrode Rather let it seat under its own weight Rotate it half a turn backward and forward to ensure the grease is spread evenly and a seal has been created Drop the cera...

Страница 8: ...n Cell Jack black Detector Electrode Jack red Turn the swivel clamp toward back of KF unit and place the entire Generator Solution Cell assembly into the Titration Cell Sleeve Turn Swivel Clamp over t...

Страница 9: ...set time navigate to time indicator and use the UP DOWN arrows to change the hour 24 hour clock and use the RIGHT LEFT arrows to move between fields Press ENTER to save values and return to Main Menu...

Страница 10: ...5 After long periods without use the unit has been turned off and or each time the reagents have been replaced you must run an initial titration to remove any ambient moisture from the reagents and c...

Страница 11: ...m the inside of the syringe The tip of the plunger should exactly align with the sample volume marking Press NEW SAMPLE key on Keypad The display will read Inject Sample and a thirty second countdown...

Страница 12: ...h sample log Data for any individual sample can be retrieved by using the UP DOWN arrows to navigate to the desired sample number and pressing ENTER The display will show the date time water percentag...

Страница 13: ...ill appear Type in the desired name for the output file in the format Filename xls and hit save Example August titrations xls A window will open with all the sample files listed Select all the KFF fil...

Страница 14: ...G WITHTHE OPTIONAL BLUETOOTH PRINTER 1 2 3 4 5 MENU Reagent Life Date Time Saved Samples Language Calibration Version Info English 78 Aug15 98 BT Mode 09 21 BT Mode Modem Printer Scanning for Printer...

Страница 15: ...ueTooth icon will appear at the top of the screen This indicates the devices have been paired successfully To print from the MAIN MENU use the navigation arrows to scroll down to Saved Samples and pre...

Страница 16: ...nother record to the printer while a record is still printing NOTE if at any time the KF unit is turned off you must turn off the printer then follow the steps on page 14 of this manual to reconnect t...

Страница 17: ...on the side of the printer two times in quick succession The printer will shut down Wait until the blue light on the printer stops blinking and turns off Then press the power button once more The pri...

Страница 18: ...om the cell Dispose of fluids according to SDS sheet s Empty contents of Generator Solution Cell by pouring into a suitable container or aspiration Dispose of fluids according to the SDS sheet Thoroug...

Страница 19: ...KF Karl Fischer Moisture Analyzer due to the potential for the other types of grease to absorb moisture and interfere with titration if it is silicon based KAM CONTROLS is not responsible for problems...

Страница 20: ...night discard the nitric acid in the cell and rinse the inside and outside of the cell with running tap water until the yellow color disappears Place distilled water in the cell and place it in a 400...

Страница 21: ...Calibration Verification Check Solution 1 1 ml and 10 l Gastight Syringes 14 and 19 gauge Needles package of rubber Septums 100 Funnels 2 Special Grease 1 and Silica Gel 1 REAGENTS and CONSUMABLES GL...

Страница 22: ...SAMPLE INJECTION PORT P 09044 STIR BAR P 09050 DRYING TUBE P 09029 RUBBER SEPTUM pack of 100 P 09015 SPECIAL GREASE 25 g tube P 370420 REPLACEMENT RECHARGEABLE BATTERY for model 09031 P 370400 POWER...

Отзывы: