Troubleshooting 

The dressing, once at pressure, is designed to need relatively little 

attention. If the pump is signaling the presence of a pressure leak 

 

(

), the following steps should be followed:

1.  Kink the tubing near the pump and wait two minutes. If the leak 

icon (

) does not disappear, the problem is in the pump and 

the clinician should be notified of the system situation/issue.

2.  Next kink the pigtail tubing near the dressing and wait two 

minutes. If the leak icon (

) does not disappear, the problem 

is in the pigtail connector and the clinician should be notified of 

the system situation/issue.

3.  If the pump and the pigtail are fine, the problem is in the 

dressing seal. If the dressing is accessible (not wrapped in a total 

contact cast or other hard to remove wrap), run your fingers 

along the gasket on the sides of the dressing. Reaffirming the 

seal should eliminate the leak within a couple of minutes. If this 

fails, the dressing is likely punctured and needs to be changed. 

Call your clinician and notify them of the situation/issue. 

NOTE:

 if the dressing cannot be reached, detach the pump and 

call your clinician for instructions

Содержание 140cc Dressing

Страница 1: ...K a l y p t o M e d i c a l NPD 1000TM Negative Pressure Wound Therapy DRESSING APPLICATION GUIDE 140cc 12x16cm Dressing...

Страница 2: ...on on how to apply the Kalypto Medical 140cc Dressing for use with the NPD 1000 NPWT System The NPD 1000 Dressing Technology is a state of the art product that delivers the benefits of negative pressu...

Страница 3: ...be used to prepare the peri wound area as this will effectively remove any remaining moisture Make sure that any hair is clipped prior to the application of the dressing A protective barrier wipe suc...

Страница 4: ...ation process but DO NOT stretch the dressing cover at the edges as you apply it NOTE As the dressing cover is a semi occlusive polyurethane stretching may induce unintended skin trauma in the peri wo...

Страница 5: ...the skin taking care to minimize folds and wrinkles while sealing the gasket for a proper air tight seal Fig 7 4 Run your finger along the perimeter of the gasket to secure the gasket material to the...

Страница 6: ...y negative pressure dressing care should be taken to secure the pressure tubing between the dressing and the pump such that it does not become a tripping hazard Running the tubing inside the patient s...

Страница 7: ...t two minutes If the leak icon does not disappear the problem is in the pigtail connector and the clinician should be notified of the system situation issue 3 If the pump and the pigtail are fine the...

Страница 8: ...ghts MN 55120 Website www kalyptomedical com Email info kalyptomedical com Phone 651 289 0284 TOLL FREE 877 286 3740 Fax 651 289 0285 P N 50037 Rev B Kalypto Medical 10 09 KALYPTO MEDICAL CONTACT INFO...

Отзывы: