Applying the Dressing 

1.  Remove the release liner from  

the dressing exposing the  

adhesive side of the dressing  

and the sealing gasket. (Fig. 4) 

 

 

 

 

2.  Carefully place the dressing over the wound, ensuring the 

absorbent pad is over the wound. (Fig. 5) Once placement 

is established, begin sealing the gasket ring, at the dressing 

bottom and top. Avoid wrinkling the dressing cover during the 

application process, but DO NOT stretch the dressing cover 

at the edges as you apply it. (

NOTE

: As the dressing cover is a 

semi-occlusive polyurethane, stretching may induce unintended 

skin trauma in the peri-wound area, such as dermatitis and/

or blistering.)  The gasket is self-sealing, so a small amount of 

wrinkling will not necessarily compromise the airtight nature of 

the dressing. Do not force the wound dressing into the wound. 

(Fig. 6) The application of negative pressure will cause the 

dressing to conform to the surface of the wound.

                 

 

        

 

Fig. 5: 

Dressing placement

 

Fig. 6:

 Securing dressing 

top & bottom gasket  

seal

Fig. 4: 

Removing 

the dressing 

release liner

Содержание 140cc Dressing

Страница 1: ...K a l y p t o M e d i c a l NPD 1000TM Negative Pressure Wound Therapy DRESSING APPLICATION GUIDE 140cc 12x16cm Dressing...

Страница 2: ...on on how to apply the Kalypto Medical 140cc Dressing for use with the NPD 1000 NPWT System The NPD 1000 Dressing Technology is a state of the art product that delivers the benefits of negative pressu...

Страница 3: ...be used to prepare the peri wound area as this will effectively remove any remaining moisture Make sure that any hair is clipped prior to the application of the dressing A protective barrier wipe suc...

Страница 4: ...ation process but DO NOT stretch the dressing cover at the edges as you apply it NOTE As the dressing cover is a semi occlusive polyurethane stretching may induce unintended skin trauma in the peri wo...

Страница 5: ...the skin taking care to minimize folds and wrinkles while sealing the gasket for a proper air tight seal Fig 7 4 Run your finger along the perimeter of the gasket to secure the gasket material to the...

Страница 6: ...y negative pressure dressing care should be taken to secure the pressure tubing between the dressing and the pump such that it does not become a tripping hazard Running the tubing inside the patient s...

Страница 7: ...t two minutes If the leak icon does not disappear the problem is in the pigtail connector and the clinician should be notified of the system situation issue 3 If the pump and the pigtail are fine the...

Страница 8: ...ghts MN 55120 Website www kalyptomedical com Email info kalyptomedical com Phone 651 289 0284 TOLL FREE 877 286 3740 Fax 651 289 0285 P N 50037 Rev B Kalypto Medical 10 09 KALYPTO MEDICAL CONTACT INFO...

Отзывы: