Optisure Remote User’s Manual
36
March 2018
APPENDIX A
Appendix A
Technical Specifications
Dew-Point Sensor Performance
Accuracy
±0.18°F (±0.1°C )
Reproducibility
±0.09°F (±0.05°C)
Measurement Technology
Chilled Mirror
Sensor
2-Stage
High Temp PEEK
Climatic head
Dew-Point Range
–40°F/Cdp @
sensor temp of
+68°F(+20°C)
+194°F(+90°C)dp
@ sensor temp of
+194°F(+90°C)
–40°F/Cdp @ sensor
temp of +68°F(+20°C)
+248F (+120°C)dp
@ sensor temp of
+248°F (120°C)
–10°Cdp @
sensor temp of
+68°F(+20°C)
+248°F (+120°C)
dp @ sensor
temp of +248°F
(+120°C)
Temperature Range
-40 to +194°F (-40 to
+90°C )
-40 to +248°F(-40 to
+120°C )
-40 to +248°F (-40
to +120°C)
% RH Range
< 0.5 to 100%
< 0.5 to 100%
10 to 100%
Min measured dew point
@ 20°C
–40°F/C
–40°F/C
-14°F (–10°C)
Mirror Material Options
Gold plated copper (standard), gold stud, Platinum
stud**
Sensor Body Material
Options
Acetal (standard), high temperature PEEK, Anodized
aluminum**
Response Speed
1.8°F/sec (1°C/sec) plus settling time
Operating Pressure
290 psig (20 barg) standard
High pressure version: 3625 psig (250 barg) max
Remote PRT
Temperature Measurement
4 wire Pt100, 1/10 DIN class B
Accuracy
±0.18°F (±0.1°C )
Cable Length
6.6ft (2m) (820ft (250m) max)
Optional Remote Pressure Sensor
Measurement Range
0 to 377 psia (0 to 25 bara)
Accuracy
0.25% Full Scale
**Recommended for special applications only. Consult Kahn Instruments before ordering.
Содержание Optisure Remote
Страница 1: ...Optisure Remote Chilled Mirror Hygrometer User s Manual March 2018...
Страница 2: ...Optisure Remote...
Страница 7: ...Optisure Remote User s Manual Kahn Instruments vii NOTES...
Страница 42: ...Optisure Remote User s Manual Kahn Instruments 35 APPENDIX A Appendix A Technical Specifications...
Страница 46: ...Optisure Remote User s Manual Kahn Instruments 39 APPENDIX B Appendix B...
Страница 53: ...Optisure Remote User s Manual 46 March 2018 Optisure Series Variables Editor Screen...
Страница 54: ...http www kahn com...