Optisure Remote User’s Manual
24
March 2018
OPERATION
The basic considerations for each measurement type are as follows:
In-Situ
1.
Dew Point Sensor position – will the sensor see an area of the environment
that is representative of what you want to measure? For example; you are
looking to measure the Relative Humidity of a room which is controlled by
an HVAC vent at either end (see figure – Room Measurement Example)
you will get very different readings depending on whether the sensor
is positioned at point A, point B or point C. Point C provides the most
representative sampling point given that it won’t be disturbed by the vent
or the door.
A
B
C
1
2
3
4
1. Door
2.
HVAC Duct, air into room
3, 4.
Ceiling Vents
Figure 20
Room Measurement Example
2.
Gas speed – if you are planning on installing the sensor in a duct, consider
how fast the sample gas is moving through it. Excessive flow speed will
cause displacement of the condensate layer on the mirror, leading to
unstable measurement.
3.
If this is the case, then a guard fitted over the sensor can mitigate the effects
of excessive gas speed by dissipating the sample throughout its’ surface
area. An appropriate guard can be purchased from Kahn Instruments.
Содержание Optisure Remote
Страница 1: ...Optisure Remote Chilled Mirror Hygrometer User s Manual March 2018...
Страница 2: ...Optisure Remote...
Страница 7: ...Optisure Remote User s Manual Kahn Instruments vii NOTES...
Страница 42: ...Optisure Remote User s Manual Kahn Instruments 35 APPENDIX A Appendix A Technical Specifications...
Страница 46: ...Optisure Remote User s Manual Kahn Instruments 39 APPENDIX B Appendix B...
Страница 53: ...Optisure Remote User s Manual 46 March 2018 Optisure Series Variables Editor Screen...
Страница 54: ...http www kahn com...