Optidew User’s Manual
42
October 2019
APPENDIX A
5
MAINTENANCE
5.1
Mirror Cleaning
Throughout the life of the instrument, periodic cleaning of the mirror surface and optics
window may be required. The frequency of this depends upon operating conditions and
the potential in the application for contaminants to be deposited on the mirror.
The Optidew will notify the user on the state of mirror contamination. The instrument will
initially give a warning in the sensor status display (or as a Magenta/flashing Magenta
indication on the transmitter version) when contamination is detected, but will continue
to operate. Cleaning the mirror, then running a DCC is necessary when this warning is
displayed.
If the contamination reaches levels which will drastically affect performance, a fault
alarm will trip, causing the instrument to switch to standby mode until action is taken.
For remote indication of an optics warning, the process alarm contact can be set to trip
whenever the optics warning is active. See sections 3.2 and 4.6 for further information.
Figure 31
Mirror contamination warning
The cleaning procedure is as follows:
1.
Set the instrument to Standby
2.
If mounted in a sample block, disconnect the sensor cable and remove
the sensor from the block.
3.
Clean the mirror surface and optics window firstly with a cotton bud/Q-
Tip soaked in distilled water, then with one of the following solvents:
methanol, ethanol, or isopropyl alcohol. To avoid damage to the mirror
surface do not press too firmly on the cotton bud/Q-Tip when cleaning.
Allow the cleaning solvent to fully evaporate.
4.
Press the ‘Calibrate Optics’ button in the ‘Alarms’ screen. For the Optidew
Wall Mount without display, refer to section 3.3.1.
Содержание Optidew Bench Top
Страница 2: ...Optidew For Kahn Instruments contact information please go to www kahn com...
Страница 13: ...Optidew User s Manual 4 October 2019 INSTALLATION This page is intentially left blank...
Страница 53: ...Optidew User s Manual 44 October 2019 APPENDIX A Appendix A Technical Specifications...
Страница 56: ...Optidew User s Manual Kahn Instruments 47 APPENDIX B Appendix B Modbus Register Map...
Страница 65: ...Optidew User s Manual 56 October 2019 APPENDIX C Appendix C Dimensional Drawings...
Страница 70: ...Optidew User s Manual Kahn Instruments 61 APPENDIX C Dual Stage Dewpoint Sensor...
Страница 71: ...Optidew User s Manual 62 October 2019 APPENDIX C Sample Block...
Страница 74: ...Optidew User s Manual Kahn Instruments 65 APPENDIX C...
Страница 75: ...Optidew User s Manual 66 October 2019 APPENDIX C...
Страница 76: ...Optidew User s Manual Kahn Instruments 67 APPENDIX C...
Страница 77: ...Optidew User s Manual 68 October 2019 APPENDIX C...
Страница 78: ...Optidew User s Manual Kahn Instruments 69 APPENDIX C...
Страница 79: ...Optidew User s Manual 70 October 2019 APPENDIX C...
Страница 80: ...Optidew User s Manual Kahn Instruments 71 APPENDIX C...
Страница 81: ...Optidew User s Manual 72 October 2019 APPENDIX C...
Страница 82: ...Optidew User s Manual Kahn Instruments 73 APPENDIX C...
Страница 83: ...Optidew User s Manual 74 October 2019 APPENDIX C...
Страница 84: ...Optidew User s Manual Kahn Instruments 75 APPENDIX C...
Страница 85: ...Optidew User s Manual 76 October 2019 APPENDIX C...
Страница 86: ...Optidew User s Manual Kahn Instruments 77 APPENDIX C...
Страница 87: ...Optidew User s Manual 78 October 2019 APPENDIX C...
Страница 88: ...Optidew User s Manual Kahn Instruments 79 APPENDIX C...