Optidew User’s Manual
34
October 2019
OPERATION
4.3
Good Measurement Practice
4.3.1 Sampling Hints
Measurement of moisture content is a complex subject, but does not need to be difficult.
This section aims to explain the common mistakes made in measurement situations, the
causes of the problem, and how to avoid them. Mistakes and bad practices can cause
the measurement to vary from the expectation; therefore a good sampling technique is
crucial for accurate and reliable results.
Transpiration and Sampling Materials
- 20
- 30
- 40
- 50
- 60
- 70
1
2
3
4
5
Time (hours)
Dew
point
(º
C)
nylon
copper
polyethylene
PTFE
nickel
stainless steel
Figure 27
Material permeability comparison
All materials are permeable to water vapor, as the water molecule is extremely small
compared to the structure of solids, even when compared to the crystalline structure
of metals. The graph above shows the dewpoint inside tubing of different materials
when purged with very dry gas, where the exterior of the tubing is in the ambient
environment.
Many materials contain moisture as part of their structure, particularly organic materials
(natural or synthetic), salts (or anything which contains them) and anything which
has small pores. It is important to ensure that the materials used are suitable for the
application.
If the partial water vapor pressure exerted on the outside of a compressed air line is
higher than on the inside, the atmospheric water vapor will naturally push through the
porous medium causing water to migrate into the pressurized air line. This effect is
called transpiration.
Adsorption and Desorption
Adsorption is the adhesion of atoms, ions, or molecules from a gas, liquid, or dissolved
solid to the surface of a material, creating a film. The rate of adsorption is increased at
higher pressures and lower temperatures.
Desorption is the release of a substance from or through the surface of a material. In
constant environmental conditions, an adsorbed substance will remain on a surface
Содержание Optidew Bench Top
Страница 2: ...Optidew For Kahn Instruments contact information please go to www kahn com...
Страница 13: ...Optidew User s Manual 4 October 2019 INSTALLATION This page is intentially left blank...
Страница 53: ...Optidew User s Manual 44 October 2019 APPENDIX A Appendix A Technical Specifications...
Страница 56: ...Optidew User s Manual Kahn Instruments 47 APPENDIX B Appendix B Modbus Register Map...
Страница 65: ...Optidew User s Manual 56 October 2019 APPENDIX C Appendix C Dimensional Drawings...
Страница 70: ...Optidew User s Manual Kahn Instruments 61 APPENDIX C Dual Stage Dewpoint Sensor...
Страница 71: ...Optidew User s Manual 62 October 2019 APPENDIX C Sample Block...
Страница 74: ...Optidew User s Manual Kahn Instruments 65 APPENDIX C...
Страница 75: ...Optidew User s Manual 66 October 2019 APPENDIX C...
Страница 76: ...Optidew User s Manual Kahn Instruments 67 APPENDIX C...
Страница 77: ...Optidew User s Manual 68 October 2019 APPENDIX C...
Страница 78: ...Optidew User s Manual Kahn Instruments 69 APPENDIX C...
Страница 79: ...Optidew User s Manual 70 October 2019 APPENDIX C...
Страница 80: ...Optidew User s Manual Kahn Instruments 71 APPENDIX C...
Страница 81: ...Optidew User s Manual 72 October 2019 APPENDIX C...
Страница 82: ...Optidew User s Manual Kahn Instruments 73 APPENDIX C...
Страница 83: ...Optidew User s Manual 74 October 2019 APPENDIX C...
Страница 84: ...Optidew User s Manual Kahn Instruments 75 APPENDIX C...
Страница 85: ...Optidew User s Manual 76 October 2019 APPENDIX C...
Страница 86: ...Optidew User s Manual Kahn Instruments 77 APPENDIX C...
Страница 87: ...Optidew User s Manual 78 October 2019 APPENDIX C...
Страница 88: ...Optidew User s Manual Kahn Instruments 79 APPENDIX C...