
Optidew User’s Manual
Kahn Instruments
13
OPERATION
2.4
Sensor Installation
The dewpoint sensor contains the optical system and the chilled mirror. It is equipped
with a 12-pin M12 connector to allow easy and secure connection to the instrument
using the supplied sensor cable.
The available options for sensor installation are:
•
via a permanently installed sample port into which the remote sensor can
be inserted or
•
via a sensor block immediately attached to the sensor around which the
sample circulates or
•
in an ambient environment where the sample is diffusing through the
sensor.
NOTE: Insure that the mirror surface is cleaned before installation. See
Section 5 (Maintenance) for cleaning details.
Connect the remote sensor cable to the sensor and to the instrument via the connector
on the rear panel. The connector is a standard M12. Align the locating pin with the slot
on the socket and press the connector into place. Rotate the outer collar of the cable-
mounted part in a clockwise direction until finger tight.
If exchanging the sensor, refer to section 5.2.
2.4.1 Environmental monitoring
If the instrument is to monitor the conditions in an environment, the sensor must be
located in a representative position, i.e. not under an air conditioning vent.
A sensor wall mounting bracket is available to conveniently secure the sensor to a wall
or panel.
NOTE: It is recommended that the sensor is equipped with the porous
aluminium guard to baffle it from flowing air currents.
2.4.2 Sample flow monitoring
If the sensor is installed within a sealed gas system it must be fixed securely without any
possibility of leaks. Insure that the sample flow across the sensor is correctly regulated.
The gas connections for the remote sensor are either via a permanently installed sample
port into which the remote sensor can be inserted or via a sensor block immediately
attached to the sensor around which the sample circulates. Gas sample entry into the
sensor block is via couplings that can be installed into the provided ⅛” NPT female
threads. A bonded seal is provided to fill the connection between the sensor and the
block.
Insure that all connections to and from the sensor block are made with appropriate
materials and fittings for moisture measurement. For guidance on suitable apparatus,
see section 4.3.
Содержание Optidew Bench Top
Страница 2: ...Optidew For Kahn Instruments contact information please go to www kahn com...
Страница 13: ...Optidew User s Manual 4 October 2019 INSTALLATION This page is intentially left blank...
Страница 53: ...Optidew User s Manual 44 October 2019 APPENDIX A Appendix A Technical Specifications...
Страница 56: ...Optidew User s Manual Kahn Instruments 47 APPENDIX B Appendix B Modbus Register Map...
Страница 65: ...Optidew User s Manual 56 October 2019 APPENDIX C Appendix C Dimensional Drawings...
Страница 70: ...Optidew User s Manual Kahn Instruments 61 APPENDIX C Dual Stage Dewpoint Sensor...
Страница 71: ...Optidew User s Manual 62 October 2019 APPENDIX C Sample Block...
Страница 74: ...Optidew User s Manual Kahn Instruments 65 APPENDIX C...
Страница 75: ...Optidew User s Manual 66 October 2019 APPENDIX C...
Страница 76: ...Optidew User s Manual Kahn Instruments 67 APPENDIX C...
Страница 77: ...Optidew User s Manual 68 October 2019 APPENDIX C...
Страница 78: ...Optidew User s Manual Kahn Instruments 69 APPENDIX C...
Страница 79: ...Optidew User s Manual 70 October 2019 APPENDIX C...
Страница 80: ...Optidew User s Manual Kahn Instruments 71 APPENDIX C...
Страница 81: ...Optidew User s Manual 72 October 2019 APPENDIX C...
Страница 82: ...Optidew User s Manual Kahn Instruments 73 APPENDIX C...
Страница 83: ...Optidew User s Manual 74 October 2019 APPENDIX C...
Страница 84: ...Optidew User s Manual Kahn Instruments 75 APPENDIX C...
Страница 85: ...Optidew User s Manual 76 October 2019 APPENDIX C...
Страница 86: ...Optidew User s Manual Kahn Instruments 77 APPENDIX C...
Страница 87: ...Optidew User s Manual 78 October 2019 APPENDIX C...
Страница 88: ...Optidew User s Manual Kahn Instruments 79 APPENDIX C...