Easidew Plus Portable Hygrometer User’s Manual
Kahn Instruments
15
MAINTENANCE
5.6
Troubleshooting
Symptom
Cause
Actions
Display shows “
ErrI
”
Sensor disconnected or
sensor element / sensor
thermistor fault.
NOTE: This error is
normally displayed
during the first few
seconds while the
instrument is starting
up.
Check sensor cable is securely
connected inside instrument.
Check instrument electronics with
4-20mA source.
Exchange sensor.
Display shows “
ErrL
”
Sensor reading under-
range.
Sensor current signal
< 4mA.
Sensor element open
circuit.
Check instrument electronics with
4-20mA source.
Exchange sensor.
Display shows “
ErrH
”
Sensor reading over-
range.
Sensor current signal >
20mA.
Sensor element short
circuit.
Sensor may have been exposed
to saturation conditions or liquid
contamination. Check that filter and
sensor guard are clean and dry.
If liquid water has contacted sensor,
or sensor block, then disassemble
and dry thoroughly. After drying, it
is recommend to purge assembled
instrument with very dry (-103°Fdp
(-75°Cdp) / 1ppm
V
moisture) air for
12 to 24 hours.
Exposure to other contaminants can
cause lasting damage and may require
sensor to be exchanged.
Check instrument electronics with
4-20mA source.
Display flickers on
and off
Battery voltage low.
Connect charger.
Display off but
battery meter full
Battery voltage low.
Connect charger.