Easidew Plus Portable Hygrometer User’s Manual
14
97478 Issue 1, October 2014
MAINTENANCE
•
Replace the locking screw and cable gland.
5.3
Filter Cartridge Replacement
Unscrew the gas inlet fitting to reveal the filter, which can then simply be removed for
checking or replacement.
5.4
Checking the Easidew Plus Portable Hygrometer Electronics Calibration
Supply
4-20 mA
PIN 3
PIN 1
(GND)
Screen/Shield
Optional
Figure 4
2-Wire Connection - View Showing Rear of Connector Terminal Block
To verify if the display electronics are within calibration, a 4-20mA current source can
be connected in place of the sensor.
At 4 mA the display should show -60.0°Cdp ±0.5°C (-76.0°Fdp ±0.9°F)
At 20 mA the display should show -76.0°Fdp ±0.9°F (-60.0°Cdp ±0.5°C)
If the displayed values are outside of this range, please contact Kahn Instruments'
Service Department.