RS80-0200-1 Rev d (10-2010)
DCN0533
5
2.0 Installation
2.1 Mounting
After unpacking the unit, inspect it for any evidence of shipping damage. Any claims for damage due to shipping
must be filed with the carrier who handled the package(s).
• Select a mounting location for the switch and the sensor probe. Your unit may be an integral mounting option or the
remote mounting option. The integral mounting unit consists of a single enclosure that includes the electronic module
and the sensor probe. The remote mounting unit consists of two enclosures, one containing the electronic module
and the other containing the sensor probe.
• Be sure that there is sufficient clearance around the mounting position to allow for the turning radius of the switch
or remote sensor enclosure as the unit is screwed into place. Allow sufficient room above the vessel entry to be able
to insert the probe into the opening of the vessel.
• The thread size of the vessel coupling should be 3/4" NPT for most probes. Certain special applications may utilize
couplings of different sizes.
• The information included on the label should be visible. If necessary clean the label using a cloth soaked with either
water or isopropyl alcohol.
2.1.2 Switchpoint and Mounting Position
Switchpoints on the sensor depend on the mounting position. Because the Resonator requires no calibration, the
switchpoint is determined by the position of the tuning fork.
Note:
When used in extremely light liquids (liquefied gas), select the low density Setpoint adjustment.
Top Mounted
Bottom Mounted
Side Mounted
2.1.1 Flush Tank Mounting
The Resonator can be flush mounted for high or low level indication using a flange or the standard ¾” connection.
Resonator
mounted via
¾” NPT connector