Table 6 - Setpoint Adjustment While Wet
OPERATION
LED 1
LED 2
Normal
GREEN OR RED
OFF
X for over 6 seconds
OFF
OFF
S2
OFF
GREEN
GREEN or RED
OFF
5.1 Density Adjustment - Precise Density Setpoint Adjustment (RS85 only) (cont’d)
Wet Adjustment:
When the fork is “WET” in the process, place the calibration magnet on the “X” position for more than 1 but less
than 6 seconds (See Figure 13). When both LED’s turn off, move the magnet to the “S2” position for 2 seconds,
then remove. The RS85 will then return to normal operating mode with the new set point (See Figure 14). If us-
ing push button, follow figures 15 and 16. This procedure defines a wet condition and corresponding density of
the process.
Figure 13
Figure 15
Figure 14
Figure 16