IEC
61000-4-8
3 A/m
3 A/m
Power frequency
magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Guidance and manufacturer´s declaration – Electromagnetic
immunity –for equipment and systems that are not life-supporting
Guidance and manufacturer´s declaration – Electromagnetic immunity
The Infrared Thermometer JPD-FR412 is intended for use in the
electromagnetic environment specified below. The customer or the user
of the Infrared Thermometer JPD-FR412 should assure that it is used in
such an environment.
Immunity
test
IEC
60601
test level
Compliance
level
Electromagnetic
environment -guidance
Radiated
RF
IEC
61000-4-3
3 V/m
80 MHz
to 2.5
GHz
3 V/m
Portable and mobile RF
communications
equipment should be used
no closer to any part of the
JPD-FR412,including
cables,
than
the
recommended
separation
distance calculated from the
equation applicable to the
frequency
of
the