EN
28
used in such an environment.
Emissions
test
Compliance
Electromagnetic
environment
-
guidance
RF
emissions
CISPR 11
Group 1
The Infrared Thermometer JPD-FR412
uses RF energy only for its internal
function. Therefore, its RF emissions
are very low and are not likely to cause
any interference in nearby electronic
equipment.
RF
emissions
CISPR 11
Class B
The Infrared Thermometer JPD-FR412
is suitable for use in all establishments,
including domestic establishments and
those directly connected to the public
low-voltage power supply network that
supplies buildings used for domestic
purposes.
Guidance and manufacturer's declaration – Electromagnetic
immunity –for all equipment and systems
Guidance and manufacturer´s declaration – Electromagnetic immunity
The Infrared Thermometer JPD-FR412 is intended for use in the
electromagnetic environment specified below. The customer or the user
of the Infrared Thermometer JPD-FR412 should assure that it is used
in such an environment.
Immunity
test
IEC
60601
test
level
Compliance
level
Electromagnetic
environment- guidance
Electrostatic
discharge
(ESD)
IEC
±6kV
contact
±8 kV
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the