91
17 Calibrating universal inputs
17.2.2
Universal inputs calibration default settings
Which calibration default settings are available depends on the configuration
settings of the universal input.
Open the calibration default settings:
Device menu > Calibration > Select Universal Input >
Calibration default settings
Calibration default settings for the individual operation modes
•
linear scaling
The calibration default settings enable the calibration routines to be accessed in
the particular calibration menu.
Calibration routines that are not enabled are not visible in the calibration menu.
The "linear scaling" operation mode permits adjustment of the zero point, slope,
and two-point calibration.
•
pH value measurement
The calibration default settings correspond to those for the zero-point and two-
point calibration for pH/redox/NH
3
analysis inputs.
⇨
"pH calibration default settings", page 57
•
Conductivity measurement
The calibration default settings correspond to those for the Ci calibration.
⇨
Chapter 16.2.2 "Calibration presets for Ci conductivity sensors", page 80
Sample screen:
Universal input calibra-
tion default settings in
the "linear scaling" op-
eration mode
Содержание AQUIS touch S
Страница 8: ...Inhalt ...
Страница 20: ...10 Configuration 20 Chapter 10 12 Calibration timer page 46 ...
Страница 68: ...13 Calibrating redox sensors 68 ...
Страница 78: ...15 Calibrating CR conductivity sensors 78 ...
Страница 88: ...16 Calibrating Ci conductivity sensors 88 ...
Страница 96: ...17 Calibrating universal inputs 96 ...
Страница 106: ...18 Calibrating O DO sensors 106 ...
Страница 146: ...21 Annex 146 ...
Страница 147: ......
Страница 148: ......
Страница 149: ......