20 Technical data
120
20.9
Voltage supply outputs, power supply unit board
20.10 Voltage supply outputs, optional board
l
Description
Output voltage
Total ampaci-
ty
a
Connection
PWR OUT
AC 110 to 240 V +10/-15%;
48 to 63 Hz or
AC/DC 20 to 30 V; 48 to 63 Hz
4 A
Spring-cage ter-
minals
a
The sum total of the output currents for the two PWR OUT connections must
not exceed the total ampacity.
Description
Output voltage
Ampacity
Connection
DC 24 V voltage supply
for
external transmitters
a
DC 24 V +5 / -5 %
42 mA
screw terminals
DC
±
5 V voltage supply
(e.g. for ISFET pH-sen-
sors)
DC +5 V +10 / 0 %
(between terminals 3 and 4)
150 mA
DC -5 V ±15 %
(between terminals 5 and 4)
30 mA
a
An optional board for voltage supply outputs accommodates all the outputs
listed in this table. A maximum of 1 such optional board can be integrated into a
device.