47
10 Configuration
10.13 Serial interfaces
The serial interface settings of all participating devices on a bus
must match.
Open:
Device Menu > Configuration > Serial Interfaces >
Serial Interfaces 1 to 2
Configuration point Selection/
setting option
Explanation
Protocol
Modbus slave
Modbus digital sensors
Communication protocol of the bus system
Modbus Slave:
For operation of the device as
a slave in a Modbus system
Modbus digital sensors:
For operation of
digiLine sensors on the serial interface (see
order details for extra code "JUMO digiLine
protocol activated")
In the JUMO AQUIS touch P, either the inter-
face on the base unit or the optional serial in-
terface (if present) can be configured for
digital sensors (JUMO digiLine operation). Si-
multaneous use of both interfaces for JUMO
digiLine operation is not possible.
Baud rate
9600
19200
38400
Transmission speed (symbol rate) of the serial
interface
a
For all bus users (device and digital sensors)
to communicate with one another, they must
have the same baud rate.
The baud rates of JUMO tecLine and JUMO
digiLine sensors are automatically set by the
JUMO digiLine master-device during the
scanning process.
When connecting digital
JUMO ecoLine sen-
sors
, the
baud rate
of the JUMO AQUIS
touch S must be set to
"9600"
prior to startup.
Otherwise, the sensors will not begin opera-
tion.
supported baud rates of JUMO digital sen-
sors
•
JUMO ecoLine: 9600
•
JUMO tecLine and JUMO digiLine:
9600,19200, 38400
Содержание AQUIS touch S
Страница 8: ...Inhalt ...
Страница 20: ...10 Configuration 20 Chapter 10 12 Calibration timer page 46 ...
Страница 68: ...13 Calibrating redox sensors 68 ...
Страница 78: ...15 Calibrating CR conductivity sensors 78 ...
Страница 88: ...16 Calibrating Ci conductivity sensors 88 ...
Страница 96: ...17 Calibrating universal inputs 96 ...
Страница 106: ...18 Calibrating O DO sensors 106 ...
Страница 146: ...21 Annex 146 ...
Страница 147: ......
Страница 148: ......
Страница 149: ......