MV / MW
9
5.4.
Countercooling
25a
25b
25b
25a
For applications near the ambient temperature, the
cooling coil must be connected to the water mains.
Using tubing, connect the cooling coil (25a) to the tap
water supply, and lead the tap water in a sink through
the return connector (25b).
A specific water flow rate of 130 ml/minute (MW) or
100 ml/minute (MV) is sufficient to compensate for the
characteristic temperature.
By installing the MVS controller (26) and a solenoid
valve (27), sudden temperature increases can be rapidly
reversed by tap water flowing through the built-in
cooling coil.
Low water consumption: For applications requiring
constant near-ambient temperature, tap water is only
used in small amounts. Configuration see page 22.
Order No.
Description
9 790 000
MVS controller
8 980 700
Solenoid valve, 220 V
26
27
25b
25a