MV / MW
7
5.
Preparations
5.1.
Installation
!
Heating circulators
Place the circulator in an upright position.
Bridge mounted circulator
Place the bath tank in an upright position.
Adjust the bridge to width of the bath tank and
place it on the bath tank.
Bridge is extendable from 310 mm to 660 mm.
5.2.
Bath liquids and tubing
Operation with
flammable
bath liquids permitted.
Recommended bath liquids:
Bath liquids
Temperature range
Flash point
fire point
Thermal M
+50 °C ... 170 °C
>275 °C
>320 °C
Thermal H
+50 °C ... 250 °C
>280 °C
>350 °C
deionized water
5 °C bis 80 °C
ATTENTION:
The maximum permissible viscosity is 30 mm
2
x s
-1.
No liability for use of other bath liquids!
Order No.
Bath liquid
8 940 100
8 940 101
8 940 102
8 940 103
Thermal M
Thermal M
Thermal H
Thermal H
10 liters container
5 liters container
10 liters container
5 liters container