Quality Management System
6
2.
Q
uality
M
anagement
S
ystem
The JULABO
Q
uality
M
anagement
S
ystem:
Development, production and distribution of temperature
application instruments for research and industries conform to
the requirements according to DIN EN ISO 9001:1994-08.
Certificate Registration No. QA 051004008.
3.
Unpacking and checking
Unpack the circulator and accessories and check for damages incurred during
transit. These should be reported to the responsible carrier, railway, or postal
authority, and a request for a damage report should be made. These instructions
must be followed fully for us to guarantee our full support of your claim for
protecting against loss from concealed damage. The form required for filing such a
claim will be provided by the carrier.
4.
Description
The JULABO circulators are suitable for temperature application to liquids in a
bath tank.
The main functional elements are the heater, circulation pump, and control
electronics. An electronic proportional temperature control (PID characteristic)
adapts the heat supplied to the thermal requirements of the bath.
Setting is rapid and simple using the keypad with its easy to learn symbols. Keypad
is splash-proof, easily cleaned and ergonomically designed.
The microprocessor technology allows four temperature values to be stored and
indicated on the MULTI-DISPLAY (LED): working temperatures T1 and T2, high
and low temperature warning limits.
The safety value for excess temperature protection, a safety installation
independent from the control circuit, is adjustable on the front and visible on the
MULTI-DISPLAY (LED).
The RS232C port permits modern process engineering without additional interface,
directly on-line, from the circulator to your application equipment.
The circulator conforms to the safety requirements specified by DIN 12 876
(safety class III), as well as DIN 58 966, and the guideline for first voltage range
EN 61010.